E.3.2 Checking/replacing the particle filter
Checking the particle filter:
Particle filters of the modular flue gas probe must
be checked regularly for contamination:
Check visually by looking through the window of
the filter chamber.
Replace the filter if there are signs of
contamination.
Replacing the particle filter:
Filter chamber may contain condensate.
1 Open the filter chamber: Turn gently
anticlockwise.
2 Remove the filter plate and replace it with a
new one (0554 3385).
3 Fit the filter chamber and lock it: Turn gently clock-
wise.
E.4 Basic operating steps
E.4.1 Switching the measuring instrument on
.
- The start screen is displayed (for about 5 s).
- Display illumination is switched on for 10 s.
Option:
To go directly to a measurement while the start screen is being displayed, press
the function key for the desired measurement. See also
Start key configuration,
p. 29.
- The
Measurements
menu is opened.
-oor-
- If another probe/sensor is connected rather than a flue gas probe: the measuring
menu for that probe/sensor is opened.