6 Using the product
35
6.2.1.4. Measurement
view
Only those parameters and units that are enabled in the measured
value display appear in the measured value display, in the saved
measurement records and on the record printouts.
> Before carrying out measurements, set up the measured value
display in such a way that the required parameters and units
are enabled, see
Measurement view, page
26
.
6.2.1.5. Setting the fuel
Before carrying out measurements, the fuel must be correctly
selected, see
Fuels, page
32
.
6.2.2. Flue
gas
To achieve usable measurement results, the measurement
period of a flue gas measurement should be approx. 3 min
and the measuring instrument should display stable
measuring values.
Call up function:
1.
[ ]
→
Measurement options
→
[OK]
→
Flue Gas
→
[OK]
.
2. Select the fuel
→
[OK]
.
Carrying out the measurement:
1. Start measurement:
[
]
.
If a draught measurement has already been carried out
separately, the value obtained is applied.
- The measured values are displayed.
2. End measurement:
[
]
.
Options
>
[Options]
→
Clipboard
: Data is saved to the clipboard.
>
[Options]
→
Delete clipboard
: Any data saved to the clipboard
is deleted.
>
[Options]
→
Save
: The measured values are saved in a
record.
>
[Options]
→
Number of lines
: Change the number of
measured values per display page.
>
[Options]
→
Recalibrate
: The gas sensors are set to zero.
>
[Options]
→
Measurement view
: (This function is not available
during a measurement): The measured value display menu is
opened.