Chapter 3 – Operation
TestEquity 123C Temperature Chamber
Page 3-1
Chapter 3 – Operation
Introduction
The Front Panel Switches control power to the chamber. The Front Panel Lights provide
indication of heat and cool functions.
The EZ-Zone Limit Controller is a protection device. It turns the chamber OFF if the workspace
temperature exceeds either a high temperature or low temperature limit set point.
The Temperature Controller controls the temperature of the chamber. The Temperature
Controller automatically turns the refrigeration system on or off as required based on the
deviation from temperature set point.
Refer to the separate F4T or F4 Temperature Controller manual for details on how to use
the Temperature Controller.
Summary of Chamber Operation
1.
Turn the Power Switch ON.
2.
Enter the appropriate high and low temperature safety limits on the Limit Controller.
3.
Enter the desired set points (or program) on the Temperature Controller.
4.
Load your test sample in the chamber.
5.
Turn the CONDITIONING Switch ON. Alternatively, turn the CONDITIONING Switch to
the EVENT 1 position and turn the Power button on the F4T controller or EVENT 1 on the
F4 controller ON.