Chapter 8 – Maintenance
TestEquity 123C Temperature Chamber
Page 8-1
Chapter 8 – Maintenance
!
WARNING:
Maintenance must be performed by properly trained personnel only.
Preventive Maintenance Intervals
Daily or As Needed
•
Clean chamber interior and exterior.
•
Listen for abnormal noise or vibration.
Every 3 Months
•
Inspect the door seal.
•
Inspect the refrigeration machinery compartment.
•
Verify the chamber performance.
Every 6 Months
•
Inspect the electrical compartment.
•
Clean the condenser.
Every 12 Months
•
Verify the calibration.