Chapter 2 – Installation 

 

 

Page 2-2 

TestEquity 123C Temperature Chamber 

Installation Location 

 

 

The chamber will produce a significant amount of heat during normal operation. Locate the 

chamber in a room with adequate ventilation to prevent excessive heat build-up. The chamber 

generates a heat load of up to 5,000 btuh during a continuous cool down from a high 

temperature. 

 

The chamber must be on a solid and level floor. Allow enough space around the chamber to 

permit serviceability and the removal of the service access panels, which are located on each side 

and the rear. 

 

!

 

CAUTION:

  The minimum clearance you should allow for proper ventilation around the 

chamber must be at least 6" from both the left and right side, and 12" from the 

rear.

 

 

!

 

CAUTION:

  This chamber is designed for operation in a conditioned laboratory 

environment. Operation above 30°C (85°F) or below 16°C (60°F) ambient 

room temperature is NOT recommended. 

 

Condensate Drain 

 

 

The chamber has a condensate drain connection on the rear of the chamber. This provides a way 

to remove condensate that may accumulate in the chamber during low-to-high temperature 

cycling or when the refrigeration system runs to maintain moderate temperatures. Any time the 

ambient air is subjected to temperatures below the dew point, moisture will condense out of the 

air. The effect is ice or frost during low temperature operation. When the chamber is heated 

above 0°C, the ice or frost will turn into water. 

 

The fitting accommodates a ½-inch male pipe thread. The chamber drain water is not under 

pressure, and is fed by gravity. Therefore, it must empty into an open floor drain. Alternatively, 

the chamber drain can empty into a condensate pump. You can purchase a condensate pump 

from suppliers such as Grainger (www.grainger.com). 

 

Under most circumstances, you will not see any water coming out of the drain. 

Содержание 123C

Страница 1: ...1 2018 Model 123C Temperature Chamber Serial Number 330218 and above Operation and Service Manual TestEquity LLC 6100 Condor Drive Moorpark CA 93021 Support 877 512 3457 Toll Free 805 480 0638 Corpora...

Страница 2: ......

Страница 3: ..._________________________________ 3 4 Internal Test Fixtures ___________________________________________________________________ 3 4 Chapter 4 Limit Controller_________________________________________...

Страница 4: ...____________8 7 Troubleshooting _________________________________________________________________8 9 Refrigeration System Charging Instructions_________________________________________8 11 R 404A High...

Страница 5: ...ard symbol 2 precedes an electric shock hazard CAUTION or WARNING statement Installation Safety Notices 2 WARNING The power cord is equipped with a NEMA 5 20P grounded polarized plug To prevent a shoc...

Страница 6: ...safe limits This could occur if your test sample exceeds the live load rating of the chamber or if the chamber s refrigeration system fails This chamber has a set of contacts that can be used to remov...

Страница 7: ...e the retainer Keep in mind that there is also Velcro that holds the retainer in place Place the retainer in the front of the pallet on the floor 5 Locate the two ramps under the chamber and remove th...

Страница 8: ...nment Operation above 30 C 85 F or below 16 C 60 F ambient room temperature is NOT recommended Condensate Drain The chamber has a condensate drain connection on the rear of the chamber This provides a...

Страница 9: ...ontroller automatically turns the refrigeration system on or off as required based on the deviation from temperature set point Refer to the separate F4T or F4 Temperature Controller manual for details...

Страница 10: ...t 1 Mode When the CONDITIONING Switch is in the EVENT 1 position you can enable and disable all chamber functions through the Power button on the F4T Temperature Controller or Event 1 Digital Output 1...

Страница 11: ...cts and surfaces that are hot or cold to prevent severe burns or frostbite Protective gloves are recommended CAUTION If your test sample is energized it may be capable of raising the workspace tempera...

Страница 12: ...ing low temperature operation or water when maintaining over 0 C and cooling is required To avoid moisture condensation make sure the port plugs are inserted at all times Also avoid opening the chambe...

Страница 13: ...d is written by the manufacturer Watlow for a wide variety of applications and configurations Not all features or functions are applicable Only the capabilities of a model PM6L1AJ AAAABAA are applicab...

Страница 14: ...your actual chamber temperature set point to prevent nuisance tripping 5 Press the RESET Key to return to the Home Page 6 The upper display will show the actual chamber temperature while the lower dis...

Страница 15: ...ower to your test sample if the Limit Controller s temperature limits are exceeded The safety contacts are rated as follows Resistive 10 A 250 VAC or 10 A 28 VDC Inductive 7 A 250 VAC To access the sa...

Страница 16: ......

Страница 17: ...2 flowing GN2 Gaseous Nitrogen Installation Option TE 1239 Connect a supply of GN2 with a maximum pressure of 100 psig to the 1 4 inch FPT fitting which is designated PURGE on the rear panel Dry Air I...

Страница 18: ...adjust the flow of purge gas into the chamber The flowmeter has a scale calibrated in SCFM The flow of purge gas should be adjusted to the minimum amount required to obtain the desired drying in the c...

Страница 19: ...rating properly by following the procedure in How to verify the chamber performance How can I modify the chamber to cool faster Unfortunately there is little you can do to improve upon the designed in...

Страница 20: ...s the ability of the temperature controller to accurately display a temperature measurement when compared to a standard The controller display accuracy is 1 66 C However the total measurement accuracy...

Страница 21: ...u would like to use is qualified Can Should I put a filter in front of the condenser air inlet No TestEquity does not recommend this Just follow the maintenance procedures and clean the condenser fins...

Страница 22: ......

Страница 23: ...min 53 min ultimate 65 C 15 min 21 min 32 min 45 min 54 min ultimate Note Transition times are measured after a 2 hour soak at the start temperature with an empty chamber Does not include the effect...

Страница 24: ......

Страница 25: ...only Preventive Maintenance Intervals Daily or As Needed Clean chamber interior and exterior Listen for abnormal noise or vibration Every 3 Months Inspect the door seal Inspect the refrigeration machi...

Страница 26: ...eing able to recognize changes from normal operating noises can be a valuable way to identify problems and prevent further damage Examples of noises to be aware of include Circulator motor and fan noi...

Страница 27: ...frigerant line which could break 1 Unplug the chamber from the power source 2 Remove the side and rear panels 3 Inspect for signs of water leaks 4 Inspect for signs of refrigeration tubing abrasion 5...

Страница 28: ...er temperature should slowly increase and stabilize to 85 C It should NOT overshoot beyond 85 C by more than a few tenths of a degree and the compressors should NOT need to turn ON in order to maintai...

Страница 29: ...ar terminals and burned or excessively pitted contacts on contactors How to Clean the Condenser 1 Disconnect the chamber from the power source 2 Remove the lower front panel 3 Clean the condenser and...

Страница 30: ...hamber and adjusting the temperature controller s calibration to correct for these errors is incorrect The correct way to adjust what the temperature controller displays compared to what is measured a...

Страница 31: ...ght glass 7 to the thermostatic expansion valve 20 The thermostatic expansion valve controls the feed of liquid refrigerant to the evaporator circuit of the cascade condenser and reduces the pressure...

Страница 32: ...balance pressure when the system is off within safe limits Both the low and high stages each have a high pressure switch 8A 8B respectively which turns off the entire refrigeration system in the event...

Страница 33: ...both compressors are OFF 2 Setpoint is low than chamber temperature but F4T Compressor Output F4 DigitalOut 8 is not ON 3 If the Cool light is ON compressors are ON 1 Pressure switch PS1 or PS2 is tr...

Страница 34: ...uity and only occurs in cool mode 1 Re enter values as documented in the controller manual NEVER use the Autotune feature 2 Control parameters may need to be changed for your unique test conditions 3...

Страница 35: ...performance as outlined in How to verify the chamber performance NOTE If the F4 Controller is cycling Cool Light cycles on off the sightglass may appear 1 2 to 2 3 full This is normal R 508B Low Stage...

Страница 36: ...out of service for even one day then you should purchase critical spare parts in advance Although most parts are standard and available from a variety of local distributors some parts are either harde...

Страница 37: ...0T US DC24 SSR1 200177 1 ea Relay Solid State 10A 5VDC in Omron G3NE 210T US DC5 SSR4 6 200129 2 ea Switch Carlingswitch LRA211 RA B 125N SW1 3 200023 2 ea Switch ON OFF ON Carlingswitch RC911 RA B 0...

Страница 38: ...on Cooling Danfoss 068U202700 11 100314 1 ea Filter Drier Danfoss 023Z5048 6 34 100524 2 ea High Pressure Cutout Johnson Controls P20DB 1D 8A 8B 100341 2 ea Hot Gas Bypass Regulator Sporlan ADRI 1 4 0...

Страница 39: ...Limit Sides both Alt high High only or LoW Low only L hY Limit Hysteresis 2 Change not recommended A ty Alarm Type oFF Do Not Change PAr1 Upper Display ACPu Alt none PAr2 Lower Display Lst Alt Lh s Hi...

Страница 40: ......

Страница 41: ...bor if they initiate service on their own without prior approval from TestEquity Replacement parts must be supplied by TestEquity for in warranty claims Purchaser will not be reimbursed for parts they...

Страница 42: ......

Страница 43: ...Chapter 10 Drawings TestEquity 123C Temperature Chamber Page 10 1 Chapter 10 Drawings...

Страница 44: ......

Страница 45: ...Electrical Subpanel Component Location...

Отзывы:

Похожие инструкции для 123C