“TestAir” LLC
620100, Yekaterinburg, p/o box 760
Tel. (343) 251-93-46, 8-912-26-25-616
[email protected]
www.testair.ru
Страница 1: ... TestAir LLC Yekaterinburg 2009 MICROMANOMETER WITH STATIC AND DYNAMIC PRESSURE RECEIVER KPDM 1 Users guide NAS 0000 004 UG ...
Страница 2: ...ARTING PROCEDURE 8 4 WORKING SEQUENCE 8 5 INSTRUCTIONS ON VERIFICATION 10 6 MAINTENANCE AND REPAIR 11 7 STORAGE RULES 11 8 TRANSPORTATION 12 9 MANUFACTURER S GUARANTEES 12 10 RECLAMATIONS 12 11 UTILIZATION 12 12 ACCEPTANCE AND PACKING CERTIFICATE 13 ATTACHEMENT 1 14 ATTACHEMENT 2 15 ...
Страница 3: ...acturer s guarantees values of basic parameters and specifications information on the construction design security measures in the device opera tion working sequence information on verification acceptance and utilization procedure of the de vice UG is intended for the device users The device does not release harmful substances polluting air and atmosphere has not harmful influence on the environme...
Страница 4: ...h a set of accessory primary converters for temperature measurement t of both gases and other environ ments 1 2 According to resistance to thermal effects and humidity of ambient air the device shall correspond to V3 augmentation group by GOST12997 when working in the range from 0 up to plus 50 ºC and humidity that is not more than 98 at 30 ºC and lower temperatures without humidity condensation ...
Страница 5: ...ed determination of air and gas flows m s 0 1 0 05v Limit of permissible absolute basic error of temperature measure ment ºС 1 0 02t Limit of permissible additional error due to the temperature devia tion of the device body frame from the normal one for every 10 ºС 0 5δ The device power supply is performed from a dry element with voltage I 9 Consumption current mA 12 Overall dimensions primary con...
Страница 6: ...tube of Pito Prandlia type with a di ameter of 6 mm and primary signal conversion unit micromanometer placed in a cylindrical body frame rigidly connected with the sensor through a pneumoconnector and performing a function of a handle with the help of which a sensor is held in the required position 2 3 4 Display and control unit is connected to the primary converter with a flexible four core elect...
Страница 7: ...l unit Digital display Button I Button P Toggle switch of the device switch ing on Flexible electric cable Cable connector Primary pneumoconverter Pneumojumper Sensing element sensor of a prima ry converter Primary converter for temperature meas urement of gas and liquid environments ...
Страница 8: ... 3 1 6 When measuring temperature the pneumometric converter should be disconnected from the display and control unit and TM should be connected When performing measurements a remote end of a thermal converter clad 6 mm in diameter should not be heat upper the tempera ture 300 ºС 3 1 7 Changing sensors PPC to PCT and vice versa can be performed not switching off the device sensor type is determine...
Страница 9: ...isplay with this the differential pressure values p mm w g indicated on a display are converted to the speed values v m s in accordance with the ratio ___ V 4 04 kr where k pneumometric tube coefficient recorded in the device memory After releasing P button the indication of the differential pressure values is restored Note This operation is valid for calculation of gas flow speed in the magnitude...
Страница 10: ... a plus sign 4 3 Working sequence in a temperature measurement mode 4 3 1 Switch on the device put PCT in a place where temperature should be measured wait for the steady state values on a display and read the temperature values 4 4 When pushing P button once click the values indicated on a display are copied into the device memory 4 5 When pushing 0 and P buttons simultaneously in the mode of pre...
Страница 11: ...disconnecting the sensor from the primary converter There is a capacity hermetically closed by a screw gag in the lower part of the sensor applied for collection of garbage and moisture penetrating into the central pneumochannel It can be periodically opened and cleaned in any way Here an additional filter 5 10 mg of any wool can be also placed If the liquid gets into the central channel it is nec...
Страница 12: ... from the date of the device shipment to the consumer 10 RECLAMATIONS 10 1 In the event of failure of the device or its defect during the guarantee period as well as detection of non completeness when delivered the consumer shall send the device with the fol lowing documents to the manufacturer application for repair list of defects 10 2 All submitted reclamations are registered by the manufacture...
Страница 13: ... storage Date of issue _____________________________________________________ Unit manager responsible for the acceptance _________________________________________________________________ Signature Full name Stamp here Packaging was performed by _______________________________________________ Signature Full name Verification officer _____________________________________________________ Signature Fu...
Страница 14: ...iver KPDM 1 serial number ________ is verified in accordance with the verification technique MP 07 221 2009 Verification results are recorded in a table Date of verificati on Verificatio n result Signature of verification officer Clarification of signature Verification mark stamping or number or verification certificate Note ...
Страница 15: ...KPDM 1 15 ATTACHEMENT 2 SCHEME OF ELECTRIC CABLE SOLDERING View on the connector from the side of slots ...
Страница 16: ......
Страница 17: ... TestAir LLC 620100 Yekaterinburg p o box 760 Tel 343 251 93 46 8 912 26 25 616 testair testair ru www testair ru ...