KPDM-1 

 

 

- 8 -

3 PRELIMINARY STARTING PROCEDURE  

3.1 General provisions. 

3.1.1 Liquid penetration into the PPC sensor holes is not allowed. 

3.1.2 When installing (replacing) the sensor the measures should be applied to ensure the 

connection hermiticity of the sensor and the PPC body frame pneumochannels. 

3.1.3 Before the beginning of work perform a test switching-on of the device and make sure 

that a resource of a battery has not been exhausted (a sign of the battery discharge in the lower 

left corner of the display is not indicated). 

3.1.4 When performing the pressure differential measurements, PPC in the gas flow should 

be oriented so that the axis of its upper part coincided with the direction of air movement, with this 

a central hole should be disclosed to meet the flow. If the sensor end is beyond the field of view of 

the operator, it is necessary to orient on the pneumojumper location. 

3.1.5  When  performing  measurements  in  high  temperature  flows  it  is  necessary  to  make 

sure that the sensor end close to the PPC body frame has not been heated upper the temperature 

50-60 º

С

, in case of necessity the sensor should be cooled down. 

3.1.6  When  measuring  temperature  the  pneumometric  converter  should  be  disconnected 

from the display and control unit and TM should be connected. When performing measurements a 

remote end of a thermal converter clad (6 mm in diameter) should not be heat upper the tempera-

ture 300 º

С

3.1.7  Changing  sensors  (PPC  to  PCT  and  vice  versa)  can  be  performed  not  switching-off 

the device, sensor type is determined automatically.  

4 WORKING SEQUENCE 

4.1 Working sequence in a mode of differential pressure measurement in air flues. 

4.1.1 Switch-on the device (move a toggle switch "on" in the upper position). 

4.1.2 Select a measuring mode if a pneumojumper is attached on both connecting branches 

the device measures dynamic pressure, if one or both connecting branches are free there is a stat-

ic pressure measurement performed. 

4.1.3 Place the upper sensor end into the gas flow observing the requirements of p. 3.1.4, 

push and release "I" button (click). In 1.5 seconds the digital information corresponding to the dif-

ferential pressure value in the place where the upper sensor end is located at the moment of push-

ing the button appears on a display and will be kept there until the next pushing of "I" button or un-

til switching-off the device.

 

The moment of measurement is accompanied by a characteristic sound 

in the body frame of the primary converter. 

 

Содержание KPDM-1

Страница 1: ... TestAir LLC Yekaterinburg 2009 MICROMANOMETER WITH STATIC AND DYNAMIC PRESSURE RECEIVER KPDM 1 Users guide NAS 0000 004 UG ...

Страница 2: ...ARTING PROCEDURE 8 4 WORKING SEQUENCE 8 5 INSTRUCTIONS ON VERIFICATION 10 6 MAINTENANCE AND REPAIR 11 7 STORAGE RULES 11 8 TRANSPORTATION 12 9 MANUFACTURER S GUARANTEES 12 10 RECLAMATIONS 12 11 UTILIZATION 12 12 ACCEPTANCE AND PACKING CERTIFICATE 13 ATTACHEMENT 1 14 ATTACHEMENT 2 15 ...

Страница 3: ...acturer s guarantees values of basic parameters and specifications information on the construction design security measures in the device opera tion working sequence information on verification acceptance and utilization procedure of the de vice UG is intended for the device users The device does not release harmful substances polluting air and atmosphere has not harmful influence on the environme...

Страница 4: ...h a set of accessory primary converters for temperature measurement t of both gases and other environ ments 1 2 According to resistance to thermal effects and humidity of ambient air the device shall correspond to V3 augmentation group by GOST12997 when working in the range from 0 up to plus 50 ºC and humidity that is not more than 98 at 30 ºC and lower temperatures without humidity condensation ...

Страница 5: ...ed determination of air and gas flows m s 0 1 0 05v Limit of permissible absolute basic error of temperature measure ment ºС 1 0 02t Limit of permissible additional error due to the temperature devia tion of the device body frame from the normal one for every 10 ºС 0 5δ The device power supply is performed from a dry element with voltage I 9 Consumption current mA 12 Overall dimensions primary con...

Страница 6: ...tube of Pito Prandlia type with a di ameter of 6 mm and primary signal conversion unit micromanometer placed in a cylindrical body frame rigidly connected with the sensor through a pneumoconnector and performing a function of a handle with the help of which a sensor is held in the required position 2 3 4 Display and control unit is connected to the primary converter with a flexible four core elect...

Страница 7: ...l unit Digital display Button I Button P Toggle switch of the device switch ing on Flexible electric cable Cable connector Primary pneumoconverter Pneumojumper Sensing element sensor of a prima ry converter Primary converter for temperature meas urement of gas and liquid environments ...

Страница 8: ... 3 1 6 When measuring temperature the pneumometric converter should be disconnected from the display and control unit and TM should be connected When performing measurements a remote end of a thermal converter clad 6 mm in diameter should not be heat upper the tempera ture 300 ºС 3 1 7 Changing sensors PPC to PCT and vice versa can be performed not switching off the device sensor type is determine...

Страница 9: ...isplay with this the differential pressure values p mm w g indicated on a display are converted to the speed values v m s in accordance with the ratio ___ V 4 04 kr where k pneumometric tube coefficient recorded in the device memory After releasing P button the indication of the differential pressure values is restored Note This operation is valid for calculation of gas flow speed in the magnitude...

Страница 10: ... a plus sign 4 3 Working sequence in a temperature measurement mode 4 3 1 Switch on the device put PCT in a place where temperature should be measured wait for the steady state values on a display and read the temperature values 4 4 When pushing P button once click the values indicated on a display are copied into the device memory 4 5 When pushing 0 and P buttons simultaneously in the mode of pre...

Страница 11: ...disconnecting the sensor from the primary converter There is a capacity hermetically closed by a screw gag in the lower part of the sensor applied for collection of garbage and moisture penetrating into the central pneumochannel It can be periodically opened and cleaned in any way Here an additional filter 5 10 mg of any wool can be also placed If the liquid gets into the central channel it is nec...

Страница 12: ... from the date of the device shipment to the consumer 10 RECLAMATIONS 10 1 In the event of failure of the device or its defect during the guarantee period as well as detection of non completeness when delivered the consumer shall send the device with the fol lowing documents to the manufacturer application for repair list of defects 10 2 All submitted reclamations are registered by the manufacture...

Страница 13: ... storage Date of issue _____________________________________________________ Unit manager responsible for the acceptance _________________________________________________________________ Signature Full name Stamp here Packaging was performed by _______________________________________________ Signature Full name Verification officer _____________________________________________________ Signature Fu...

Страница 14: ...iver KPDM 1 serial number ________ is verified in accordance with the verification technique MP 07 221 2009 Verification results are recorded in a table Date of verificati on Verificatio n result Signature of verification officer Clarification of signature Verification mark stamping or number or verification certificate Note ...

Страница 15: ...KPDM 1 15 ATTACHEMENT 2 SCHEME OF ELECTRIC CABLE SOLDERING View on the connector from the side of slots ...

Страница 16: ......

Страница 17: ... TestAir LLC 620100 Yekaterinburg p o box 760 Tel 343 251 93 46 8 912 26 25 616 testair testair ru www testair ru ...

Отзывы: