The EXT indicator will be lit when the manual
trigger is active.
2. Manually
Triggering
Measurements
Press the MANUAL TRIG
INT
key
each time you want to start a single
measurement.
The EXT indicator will be lit when the manual
trigger is active.
Internal Triggering
Press SHIFT + MANUAL TRIG
INT
to return the triggering mode back
to internal mode.
The EXT indicator will become cleared from the
display.
Compare Function
Background
The compare function compares a measured value to a
“Normal” value that has an upper (HI) and lower (LO)
limit. The upper and lower limit is set as a percentage of
the Normal value.
A measured value that falls within the upper and lower
limits is considered a GO (pass), a value that falls below
the lower limits is considered LO, and a value that falls
over the upper limit is a HI.