MEASUREMENT
25
Select Measurement Rate
Background
The resistance measurement speed has 2 ranges: low and
high. Low speed is the most accurate with 7
measurements/second and a full scale of 30,000 counts.
High speed has 30 measurements/second with a full scale
3,000 counts.
The rate function is applicable to the resistance,
temperature and temperature compensation
measurements.
1. Select Rate
Press the SHIFT +
AUTO/MAN
SPEED
keys to switch
between the LO and HI rates.
LO rate
The low rate is shown with a full scale of 5 digits
in the resistance measurement mode.
HI rate
The high rate is shown with a full scale of 4 digits.
Note
For detailed specifications, please see the specifications
on page 49.
Using the Trigger Function
Background
The GOM-802 can use internal or manual triggering for
the resistance, temperature and temperature
compensation measurement modes.
External triggering is only supported with the Handler
and Scan modes.
By default the GOM-802 is in internal triggering mode.
1. Select Manual
Trigger
Press MANUAL TRIG
INT
to switch
to manual triggering mode.