background image

 

6/8 

3

 

Scope of Delivery 

 

TSRP3 WLAN Wireless and USB Radon Soil Probe 

 

Power adapter 230VAC/5VDC 

 

USB cable A-micro  

 

Antenna 

 

2 pcs of dehumidifiers 

 

Operation Manual 

 

4

 

Product Specification 

 

Product 

TSR3P WLAN Wireless and USB Radon Soil Probe 

Type symbol 

042 127 192 000 

Average measurement sensitivity 

0,06 count/hour/Bq.m-3 
(method RaA+RaC; 

15°C ÷ 30°C; rel. hum. 20% ÷ 40%)

 

Measuring range 

MDA 

 100 000 Bq/m

3

;  

MDA = 100 Bq/m

3

 per 1 hour or 20 Bq/m

3

 per 24 hours 

Measurement uncertainty 

< 25% at 300 Bq/m

3

 per 1 hour; 

< 6% at 300 Bq/m

3

 per 24 hours 

Measuring chamber capacity 

0,044 dm3 

Response rate 

< 30 minutes (RaA); < 3 hours (RaA + RaC) 

Radon records 

calculated from RaA (quicker, less sensitive) 

 

calculated form RaA + RaC (slower, more sensitive) 

Measuring relative humidity range 

10 

 90 % 

Measuring temperature range 

-

20 to + 60 °C 

 

Radio interface 

868MHz  

Max number of measuring network 
elements 

16  

Probe to terminal unit distance (RF range)  depends on number of walls and building material, up to 600 m 

in open space 

Possibility of using repeater for RF range 
extending 

yes 

Results reading interval (from probe) 

240-65535 sec (4 min - 18.2 hours) 

Records saving interval (probe) 

1- 255 minutes, default 1 hour 

Results memory capacity in probe 

4096 (150 days of 1 hours records) 

Powering 

internal rechargeble accumulator; charging via USB 

Accu life after full charging 

>1 year 

Current radon concentration results  

short-term (0,5 hour running average from RaA) 

 

long-term (24 hours running average from RaA + RaC)

 

Dimension 

Ø 

50 x 280 mm 

 
 

Содержание TSRP3

Страница 1: ...chnical Specifications Operation Manual v 1 2019 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 6 4 Product Specification 6 5 Operating Instructions 7 6 Basic Maintenance 8 7 EC Declaration of Conformity 8 8 Repairs 8 9 Warranty 8 10 Accessories 8 11 Revision History 8 ...

Страница 2: ...d product If you experience any problems with your product please contact us at TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 1 Introduction This document describes technical specifications and user operation of the TSRP3 WLAN Wireless and USB Radon Soil Probe Product was developed and manufactured in the Czech Republic All rights reserved TESLA Offer or delivery of products or services re...

Страница 3: ...dict seismic activity or earthquakes Portable probe basis is a measuring chamber with a semiconductor photodetector Radon enters the chamber by diffusion through the input filter on the bottom of probe The probe measures in autonomous and time continuous way It processes results every 2 minute intervals and from this counts short term moving average of radon concentration 1 hour moving average ave...

Страница 4: ...ntral unit memory Via connected computer to Central Unit and TERAview application on PC it is possible to download and process all data from system and configure whole system Setting and configuration of Radon Soil Probe and whole system is also managed by TERAview application on PC TERAview application version 3 11 6 and higher drivers and user manual with detail configuration description is free...

Страница 5: ...ion and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuator in central unit can be use which is hardwired with power relay Power relay switches on a fan which decreases radon concentration within an area After decreasing of radon concentration actuator receives command to switch off fan This cycle repeats depending on increasing or decr...

Страница 6: ...m RaA quicker less sensitive calculated form RaA RaC slower more sensitive Measuring relative humidity range 10 90 Measuring temperature range 20 to 60 C Radio interface 868MHz Max number of measuring network elements 16 Probe to terminal unit distance RF range depends on number of walls and building material up to 600 m in open space Possibility of using repeater for RF range extending yes Result...

Страница 7: ...livered power adapter Power adapter is connected to probe via provided USB cable In case of blackout internal accumulator ensures UPS function Configuration Setting and configuration is different according to operation way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application...

Страница 8: ...fier to probe is necessary to screw and disassemble 2 and 3 chamber from the bottom see picture in chapter2 If the relative humidity of environment is above 90 and temperature 10 C the change interval of dehumidifier is 6 weeks Humid dehumidifier is possible to use repeatedly after drying it up in oven with 210 C for 1 hour The temperature cannot be over 230 C 7 EC Declaration of Conformity EC Dec...

Отзывы: