background image

 

3/8 

 

 

Radon  Soil  Probe  is  adapted  for  insertion  into  a  bore  hole  of  minimum  diameter  50mm  and  for 

operation  in  places  with  increased  humidity  as  are  caves and  mines.  Two  probe  chambers  are  filled  with 
replaceable special dehumidifier for decreasing of humidity in measuring chamber if the probe operates in 
humidity  environment  with  relative  humidity  above  90%.  If  it  operates  in  humidity  environment  with 
relative  humidity  above  90%  the  dehumidifier  needs  to  be  replaced  or  dried.  If  the  relative  humidity  of 
environment  is  above  90% 

and  temperature  10  °C 

the  change  interval  of  dehumidifier  is  6  weeks.  The 

accumulator will last for up to 1 year after one full charging ensuring autonomous operation of the probe 
within  the  bore.  It  depends  on  frequency  of  wireless  data  downloading  from  probe  and  on  humidity 
condition also.  

The TSRP3 Radon Soil Probe is designed to continually measure radon volume activity in soil. Soil radon 

concentration changes may precede certain behavior of Earth’s crust and various elements in the bedr

ock. 

For  example,  long-term  monitoring  of  the  soil  gas  radon  concentration  can  be  used  to  predict  seismic 
activity or earthquakes. 

Portable  probe  basis  is  a  measuring  chamber  with  a  semiconductor  photodetector.  Radon  enters  the 

chamber by diffusion through the input filter on the bottom of probe. The probe measures in autonomous 
and  time  continuous  way.  It  processes  results  every  2-minute  intervals  and  from  this  counts  short-term 
moving  average  of  radon  concentration  (  1  hour  moving  average  -  average  of  30  2-minute  process 
intervals). It also counts long-term moving average of radon concentration (24 hours moving average). The 
probe saves time records of these radon concentration values including values of humidity and temperature 
within its internal memory (typically at an interval of 1 hour). Next saved value is time record of measuring 
energy spectrum (typically at an interval of 12 hours). Bottom of the probe cannot be covered. The probe 
permanently measures and saves data into internal memory if accumulator is charged. 

LEDs „

STAT

“ indicate 

current status of probe see  

´

Operation manual

´

 below.  

The  resulting  values  can  be  downloaded  continuously  during  measurement  or  at  once  at  the  end  of 

measurement.  Data  from  the  radon  probe  can  be  downloaded  to  a  PC  directly  via  USB  interface  or 
wirelessly via antenna and Central Unit. Central Unit is not included with package of probe and it is sold and 
delivered extra. 

Содержание TSRP3

Страница 1: ...chnical Specifications Operation Manual v 1 2019 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 6 4 Product Specification 6 5 Operating Instructions 7 6 Basic Maintenance 8 7 EC Declaration of Conformity 8 8 Repairs 8 9 Warranty 8 10 Accessories 8 11 Revision History 8 ...

Страница 2: ...d product If you experience any problems with your product please contact us at TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 1 Introduction This document describes technical specifications and user operation of the TSRP3 WLAN Wireless and USB Radon Soil Probe Product was developed and manufactured in the Czech Republic All rights reserved TESLA Offer or delivery of products or services re...

Страница 3: ...dict seismic activity or earthquakes Portable probe basis is a measuring chamber with a semiconductor photodetector Radon enters the chamber by diffusion through the input filter on the bottom of probe The probe measures in autonomous and time continuous way It processes results every 2 minute intervals and from this counts short term moving average of radon concentration 1 hour moving average ave...

Страница 4: ...ntral unit memory Via connected computer to Central Unit and TERAview application on PC it is possible to download and process all data from system and configure whole system Setting and configuration of Radon Soil Probe and whole system is also managed by TERAview application on PC TERAview application version 3 11 6 and higher drivers and user manual with detail configuration description is free...

Страница 5: ...ion and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuator in central unit can be use which is hardwired with power relay Power relay switches on a fan which decreases radon concentration within an area After decreasing of radon concentration actuator receives command to switch off fan This cycle repeats depending on increasing or decr...

Страница 6: ...m RaA quicker less sensitive calculated form RaA RaC slower more sensitive Measuring relative humidity range 10 90 Measuring temperature range 20 to 60 C Radio interface 868MHz Max number of measuring network elements 16 Probe to terminal unit distance RF range depends on number of walls and building material up to 600 m in open space Possibility of using repeater for RF range extending yes Result...

Страница 7: ...livered power adapter Power adapter is connected to probe via provided USB cable In case of blackout internal accumulator ensures UPS function Configuration Setting and configuration is different according to operation way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application...

Страница 8: ...fier to probe is necessary to screw and disassemble 2 and 3 chamber from the bottom see picture in chapter2 If the relative humidity of environment is above 90 and temperature 10 C the change interval of dehumidifier is 6 weeks Humid dehumidifier is possible to use repeatedly after drying it up in oven with 210 C for 1 hour The temperature cannot be over 230 C 7 EC Declaration of Conformity EC Dec...