Unit Care and Maintenance
5.2 – Normal Function Test Procedures
This section deals with “normal function” test proce dures, and includes
all steps necessary to ensure that the TI200-PFC is operating within
specified para meters prior to use. A digital multimeter (an example is
shown in Figure 5.2.1) capable of measuring dc and ac voltage and
resistance will be required to perform some of the tests. These functional
test procedures should become routine.
Figure 5.2.1 – Digital
Multimeter
Check Unit for Evidence of Damage
Check for dents, punctures, case distortion or misalignment, and cracked or loose connec tors. If no damage
is evident, proceed to the next step. If damage is evident, contact Tesla™ Industries, Inc.
Check Unit Internal Resistance (Test for Shorts)
2.
Without ac power, place the nega-
tive probe on a screw of the DC Re-
ceptacle. Move the positive probe
to the dc positive post. Multimeter
should read “OL” (Open Line).
3. Without ac power, place the nega-
tive probe on a screw of the DC Re-
ceptacle. Move the positive probe
to the dc negative post. Multimeter
should read “OL” (Open Line).
1. Set multimeter to Ohms.