34
NSG 437 ESD simulator
Check the voltage level under the following conditions:
Air-discharge
Single-discharge
Polarity: positive and negative
Voltage settings: 2, 4, 8, 15 and 30 kV
Permissible tolerance < ± 5% of set value
Check the discharge current and pulse form as follows:
Contact-discharge
Single-discharge
Polarity: positive and negative
Voltage settings: 2, 4, 8, 15 and 30 kV
Compare the measured values with the reference data in IEC / EN 61000-4-2.
These values are valid only for the discharge network that conforms to IEC / EN
61000-4-2.
Remark
Teseq offers an accredited service for this kind of work!
10.2 Exchanging the R / C network
If a network needs to be exchanged, the test has to be
stopped first, followed by a waiting time of at least 5 s
to ensure the voltage being internally discharged.
Switch the simulator off.
Open the flap under the display and rotate the pistol backwards until the
network drops out under its own weight. Take care! Catch the network in the
other hand.