11
Mains input is via a pre-wired IEC 60309 3-phase plug.
Star configured supplied with a 5 pin plug.
Delta configured supplied with a 4 pin plug
If the plug requires rewiring, brown and 2 black wires are 3-phase line, blue is
neutral when used in star (unused in delta configuration) and green/yellow for
earth.
The amplifier safety mechanism is a circuit breaker. The switch unit is supplied
from the power unit via the power breaker, located on the rear panel of the
power unit. Connect the cable from the switch unit to the power unit 7 pin
socket.
The remaining connectors are: RF input, RF output and BNC interlock. The use
of the interlock is optional, and the amplifier will operate normally with this
connector left unterminated. If an attachment is made to the BNC interlock,
ensure that the connector is a 50 ohm type and that a high quality coaxial cable
is used for connection to the switch.
Make sure that the amplifier is located in a situation with adequate ventilation.
During operation, the amplifier will draw as much as 6 kVA from the mains.
Ensure that the environment of the amplifier does not cause warm air from the
fans to be cycled back into the amplifier.