RS-232C Commands in Detail
41
DOWNWARD
A command to lower the probe down.
Type
Command
Input
DOWNWARD
Response
{ OK | ERR30 | ERR40 | ERR90 }
●
OK: The probe moves downward.
●
ERR30: Probe-downward sensor error. It occurs only when the probe is in sensor mode.
(The vertical operation of fixture is incomplete.)
●
ERR40: Status error, It occurs when the command is not executable. For example:
○
OUTWARD (handler) or UPWARD (probe) command is sent while the lid is closed.
○
DOWNWARD or UPWARD command (probe) is sent while the handler is at
outward position.
○
OUTWARD command (handler) is sent while the probe is at downward position.
●
ERR90: The command is sent while the probe is off.
5.6.6 Pass/Fail Decisions
PASS
A command that decides a pass after a test.
Once the command is sent, the PASS LED on the control box will glow green (
) and the lid will
open automatically.
If the lid is already open, only the PASS LED will be on.
Type
Command
Input
PASS
Response
{ OK | ERR50 }
●
OK : The green PASS LED gets turned on, and the lid opens.
●
ERR50: Lid-open sensor error. The lid-open sensor does not detect the cylinder in open
position.