Operating without Fixture
27
The figure above is just a recommendation, so the actual connection may be adjusted for
your convenience.
4.4 Operating without Fixture
In order to work Shield Box without fixture, both fixture modes should be set to 'Off Mode.'
Refer to
5.6.2 Handler & Probe Configuration Commands
.
If any of the fixture modes is set to 'Sensor Mode,' the lid will stay open with Fail LED on, as there is no
sensor to confirm each motion.
Also, if it is set to 'Timer Mode,' the lid operation will take certain time waiting for each fixture motion,
delaying the entire process.