9
Use of vacuum option with Titan Stabilizer Attachments:
1. Attach male luer connection of accessory vacuum tubing to the
female luer port of the vacuum control stopcock, making sure the
vacuum source is switched OFF. Attach the vacuum source connector
of the accessory vacuum tubing to the vacuum source; turn the
vacuum control stopcock OFF to the vacuum source.
2. Turn the vacuum regulator of the vacuum source on to sufficient
negative pressure to affix the Titan Stabilizer Attachment to tissue.
Due to variations in OR equipment and utilities, use the least amount
of vacuum necessary for desired product performance. The regulator
should not exceed -400 mmHg.
3. Turn the vacuum control stopcock OPEN to the vacuum source
and attachment once proper position has been obtained. Verify
attachment of the stabilizer to the tissue.
4. To release attachment from tissue, turn the vacuum control stopcock
OPEN to the atmosphere prior to removal of the stabilizer attachment.
WARNING
• Use of excessive negative vacuum pressure may cause tissue
damage.
• Place vacuum control stopcock OPEN to atmosphere and OFF to
the vacuum source prior to removal of attachment from tissue.
STORAGE
The Titan Stabilizer Attachments storage conditions:
•
A clean, dust-free location out of direct sunlight and free from risk of
high temperature, excessive humidity, and water leakage.
•
Do not store the device near chemicals.
DISPOSAL
After use, dispose of product and packaging in accordance with hospital,
administrative and/or local, state, federal and international laws and
regulations.
Содержание Titan T401231
Страница 2: ...2 This page is intentionally left blank...
Страница 4: ...4 This page is intentionally left blank...
Страница 10: ...10 This page is intentionally left blank...
Страница 16: ...16 Tato str nka je mysln ponech na pr zdn...
Страница 22: ...22 Denne side er med vilje tom...
Страница 34: ...34 T m sivu on j tetty tarkoituksella tyhj ksi...
Страница 52: ...52 Ez az oldal sz nd kosan maradt res...
Страница 58: ...58 Questa pagina lasciata intenzionalmente vuota...
Страница 64: ...64 Denne siden er blank med hensikt...
Страница 78: ...78 1 Titan...
Страница 79: ...79 1 Terumo Terumo Terumo 2 2 Titan Terumo 3...
Страница 81: ...81 Titan Flex 4 20 U Titan Flex 4 Titan Flex 5 Terumo Terumo Titan 1 2 Titan 400...
Страница 82: ...82 3 4 Titan...
Страница 84: ...84 CABG 1 Terumo Terumo Terumo 2...
Страница 85: ...85 2 Titan Terumo 3 3 Titan 360...
Страница 87: ...87 Titan 1 2 Titan 400 mmHg 3 4 Titan...
Страница 88: ...88...
Страница 106: ...106 Denna sida r avsiktligen tom...
Страница 108: ...108 CABG 1 Terumo Terumo Terumo 2...
Страница 109: ...109 2 Titan Terumo 3 3 Titan 360 2 U 4 45 45...
Страница 111: ...111 Titan 1 2 Titan 400 mmHg 3 4 Titan...
Страница 112: ...112...
Страница 126: ...126 This page is intentionally left blank...
Страница 127: ...127 This page is intentionally left blank...