3.1 Measure Voltage/Display Mode
Now it will be ask for your choose off the measure voltage.
In the display appears :
VOLTGE MODE :
AUTO
(MAN.10V) (MAN. 100V)
By pressing key (B)
„
Change
”
the display is switching between AUTO (Automatic), MAN. 10V, and MAN.
100V. Press the (A)
„
Set-Up
“
key to choose the displayed voltage mode.
After that, it will be ask for your choose off the display voltage.
DISPLAY MODE :
R>2.0xE12?
(R>2.0 T? )
By pressing key (B)
„
Change
”
the display is switching between the mathematical and technical displa y form.
Press the (A)
„
Set-Up
“
key to choose the displayed display time.
Now the unit will execute a reset, and in the display appears if timer is on, else go to point 3.2 Timer off.
T=xxxs Avg Auto
(Lap) (+10V, +100V)
PRESS START !
By pressing key (A)
„
Start
”
the next measurement will be started.
The timer decrements in 1 second steps up to Zero, then the measure value will appears in the display as
R
avg
by AVERAGE - Mode and R
lap
by Lapse
–
Mode.
In the 2nd line the air temperature and the humidity is displayed:
Ravg = 2.5xE9?
(Rlab = 2.5 G? )
AIR: xx
°
C xx%r.F
Alternates all 2 second in the 2nd line appears:
FN:xxx T=xxxs
By pressing key (A)
„
Start
“
the measure value together with the air temperature and the Humidity will be
saved in the EEPROM using the displayed File
–
No. These values can be read later with the PC - software.
3.2 Timer
If the timer is not activated, the TOM 600 directly switches on the resistance display in the on-line operation.
According to the chosen display mode and voltage mode in the display appears:
Example:
R=5.5xE10W T=OFF
AUTO
AIR: xx
°
C xx%r.F
5