![Tera TOM 600 Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/tera/tom-600/tom-600_manual_1088157002.webp)
The TERA Ohmmeter TOM-600 is suitable very well for the mobile use in industrial areas through its
compact housing and through the battery power. It can also be used stationary with a plug power supply.
Important: Don
’
t use the AC-Adapter without installed rechargeable batteries.
The TOM-600 is menu-controlled, has only two buttons and so it is very operator-friendly.
All parameters are shown at the LC - Display.
The included PC-Software is for remote controlled usage, and additionally measuring results can be
administered and processed.
The TOM-600 works with the voltage
–
current measure - principle.
To the transactions of the measurements anuses of tied-up Norm DIN EN 61340 / EOS-ESD 4.1/6.1 the
measuring time can be selected with the internal timer.
At Timer = ON the measurement values inclusive the air parameters can be stored in the internal EEPROM.
They don
’
t are deleted by switching off the unit. The values can be read out trough the PC-Interface at all
times.
Since the resistance values depend very strongly on the Humidity and the air temperature,
these influences are measured together with the resistance value.
So a reproducible measure ment is given.
2.1 Start Operation
The
TERA
Ohmmeter TOM 600 is delivered with 4 x Mignon (AA) NiMH
–
Batteries, and is directly ready
to start. For charging the batteries and stationary operation use the enclosed plug power supply only.
l
The electrodes are connected to the corresponding sockets (s. legend) and place on the probe .
l
During the measurement of very high-ohm-resistors is to be kept that no influence effect is happen on
the measure input.
l
To Switch On the unit press key
„
B
“
ON/OFF.
After switching on, first the software version is shown.
*>2.0xE12W T=OFF**
Air: xx
°
C xx%r.F
Provided the timer is
„
ON
“
, in the display appears press start. You can start the measure by pressing key
“
A
”
start. The current resistance value and the timer attitude are shown afterwards.
1. Description
2. Operating Instruction
2