Programme Settings

Display  

Mode 

Frequency/Pulse Width

A1  

Burst 

80Hz 150us

A1 Boost  

Continuous 

80Hz 150us

B2  

Burst 

150Hz 150us

B2 Boost  

Continuous 

150Hz 150us

C3 

Ramp 

 80Hz 250us

17. EMC PRECAUTIONS

Use special precautions regarding EMC according to the information 

provided below.

Guidance and manufacturer’s declaration - electromagnetic emissions

The mamaTENS is intended for use in the electromagnetic environment specified below.
The customer or the user of the mamaTENS should assure that it is used in such an environment.

Emissions test 

Compliance 

Electromagnatic environment - guidance

RF emissions 

Group 1 

The mamaTENS uses RF energy only for its internal  function. Therefore,

CISPR 11 

its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment

RF emissions 

Class B 

CISPR 11 

Harmonic emissions 

Not applicable 

IEC61000-3-2
Voltage fluctuations/ flicker 

Not applicable

emissions IEC 61000-3-3

The mamaTENS is suitable for use in domestic establishment and in
establishment directly connected to a low voltage power supply 
network which supplies buildings used for domestic purposes.

The mamaTENS is intended for use in the electromagnetic environment specified below. The 

customer or the user of the mamaTENS should assure that it is used in such an environment.

Immunity test 

IEC 60601 test level 

Compliance 

Electromagnetic

level 

environment guidance

Electrostatic discharge (ESD) 

±6 kV contact 

±6 kV contact 

Floors should be wood,

IEC 61000-4-2    

±8 kV air 

±8 kV air 

concrete or ceramic tile

If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst 

±2 kV for power supply lines 

Not applicable 

Not applicable

IEC 61000-4-4 

±1 kV for Input/output lines 

Surge      

±1 kV differential mode 

Not applicable 

Not applicable

IEC 61000-4-5  

±2 kV common mode

Voltage dips, short interruptions  <5 % U

T

 (>95 % dip in U

T

) for 0,5 cycle  Not applicable 

Not applicable

and voltage variations on 

40 % U

(60 % dip in U

T

 ) for 5 cycles

power supply input lines 

70 %  U

T

 (30 % dip in U

T

 ) for 25 cycles

IEC 61000-4-11 

<5 %  U

T

 (>95 % dip in U

T

 ) For 5s

Power frequency  (50/60 Hz) 

3 A/m  

Not applicable 

Not applicable

magnetic field
IEC 61000-4-8

NOTE U

T

 is the a.c. mains voltage prior to application of the test level.

Other portable and mobile RF communications equipment can affect 

performance. Do not use when adjacent to or stacked with other 

electrical equipment. Use of leads or pads other than those listed in 

section 18 may affect EMC performance.

mama tens instructions artwork.indd   28-29

Содержание MamaTENS

Страница 1: ...INSTRUCTIONS FOR USE READ CAREFULLY BEFORE USE...

Страница 2: ...nd try out the unit to see how it feels 6 Replace electrodes onto their liners and seal them in the re sealable bag 7 Put everything back into the pouch until labour begins 8 As soon as labour begins...

Страница 3: ...4 5 Programme C 8 5 After use 9 6 Cleaning 10 7 Cautions Warnings 10 8 Controls 13 9 Positioning of pads 16 10 Batteries 10 1 Low battery display 16 10 2 Changing the batteries 17 11 Returning your re...

Страница 4: ...AND HOW DOES IT WORK Midwives recommend maternity TENS machines for their benefits particularly in early labour as they enable mums to remain mobile and upright This aids descent and dilatation and p...

Страница 5: ...is clearly marked on and off whilst the curved green button to the right of the machine is marked with arrows to increase or decrease the intensity as desired Intensity cannot be adjusted unless the e...

Страница 6: ...not break WARNING CH1 CH2 4 2 WHEN TO START USING YOUR mamaTENS You can start using mamaTENS as soon your contractions start to become regular 4 3 PROGRAMME A When you switch the machine on it automa...

Страница 7: ...he mamaTENS To correct this change over the leads and reset the intensity NB NOTE 4 5 PROGRAMME C If you find programmes A and B do not meet your needs for pain relief there is a third programme C whi...

Страница 8: ...ened pack of electrodes is 2 years This may be affected by very high temperatures or very low humidity 6 CLEANING Clean the case and lead wires after each use by wiping with a damp cloth and a solutio...

Страница 9: ...onse Over the eyes May affect eyesight or cause headaches Across the front of the head Effect on patients who have had strokes or seizures is not known May affect sense of balance The effects of stimu...

Страница 10: ...60 in Programme C Each time the is pressed the intensity will decrease by one level Hold down the or button to scroll intensity As you increase the intensity you will experience a gentle tingling feel...

Страница 11: ...this stage using spare batteries supplied See instructions below as to how to change the batteries Upper Pads Uterus Lower Pads Perineum Birth Canal 15 16 The UPWARD POINTING TRIANGULAR YELLOW BUTTON...

Страница 12: ...we ll arrange an additional two weeks absolutely free of charge Sending your maternity TENS machine back at the end of the rental period is very easy You ll have received a pre paid envelope with your...

Страница 13: ...ility and hygiene use only genuine rectangular electrodes with integral leads For units after 1330000 we are using part number E CM50100 MTNEW For units before 1330000 we are using part number E CM501...

Страница 14: ...e world s resources and minimise any adverse effects on the environment 16 TECHNICAL INFORMATION Waveform Symmetrical Bi Phasic Rectangular Amplitude 75 mA zero to peak over 500 Ohm 50 steps in A B 60...

Страница 15: ...the electromagnetic environment specified below The customer or the user of the mamaTENS should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance Electromagn...

Страница 16: ...rmal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the mamaTENS b Over the frequency range 150 kHz to 80 MHz field strengths sho...

Страница 17: ...TensCare Ltd 9 Blenheim Road Epsom Surrey KT19 9BE UK Tel 44 0 1372 723434 www tenscare co uk Distributed by Pub Ref I MT UK Rev 1 4 3 17 Manufactured by...

Отзывы: