23
24
Programme Settings
Display
Mode
Frequency/Pulse Width
A1
Burst
80Hz 150us
A1 Boost
Continuous
80Hz 150us
B2
Burst
150Hz 150us
B2 Boost
Continuous
150Hz 150us
C3
Ramp
80Hz 250us
17. EMC PRECAUTIONS
Use special precautions regarding EMC according to the information
provided below.
Guidance and manufacturer’s declaration - electromagnetic emissions
The mamaTENS is intended for use in the electromagnetic environment specified below.
The customer or the user of the mamaTENS should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnatic environment - guidance
RF emissions
Group 1
The mamaTENS uses RF energy only for its internal function. Therefore,
CISPR 11
its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment
RF emissions
Class B
CISPR 11
Harmonic emissions
Not applicable
IEC61000-3-2
Voltage fluctuations/ flicker
Not applicable
emissions IEC 61000-3-3
The mamaTENS is suitable for use in domestic establishment and in
establishment directly connected to a low voltage power supply
network which supplies buildings used for domestic purposes.
The mamaTENS is intended for use in the electromagnetic environment specified below. The
customer or the user of the mamaTENS should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
Electromagnetic
level
environment guidance
Electrostatic discharge (ESD)
±6 kV contact
±6 kV contact
Floors should be wood,
IEC 61000-4-2
±8 kV air
±8 kV air
concrete or ceramic tile
If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst
±2 kV for power supply lines
Not applicable
Not applicable
IEC 61000-4-4
±1 kV for Input/output lines
Surge
±1 kV differential mode
Not applicable
Not applicable
IEC 61000-4-5
±2 kV common mode
Voltage dips, short interruptions <5 % U
T
(>95 % dip in U
T
) for 0,5 cycle Not applicable
Not applicable
and voltage variations on
40 % U
T
(60 % dip in U
T
) for 5 cycles
power supply input lines
70 % U
T
(30 % dip in U
T
) for 25 cycles
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
) For 5s
Power frequency (50/60 Hz)
3 A/m
Not applicable
Not applicable
magnetic field
IEC 61000-4-8
NOTE U
T
is the a.c. mains voltage prior to application of the test level.
Other portable and mobile RF communications equipment can affect
performance. Do not use when adjacent to or stacked with other
electrical equipment. Use of leads or pads other than those listed in
section 18 may affect EMC performance.