The
Flexistim
can also be operated directly from the charger. Just plug it directly into the
socket in the side of the unit. This automatically disconnects the battery.
USE OF OTHER CHARGERS COULD BE HAZARDOUS AND WILL NEGATE
THE GUARANTEE
Warning
There is a risk of smoke, fire, or rupture if the battery is not used according to the following
guidelines:
• Do not disassemble the battery
• Do not short-circuit the battery
• Do not incinerate or heat the battery
• Do not use or leave battery near a fire, stove or heated place (more than 80°C)
• Do not immerse the battery in water or sea water, or get it wet
• Do not charge battery nearby the fire or in strong sunlight
• Only use the charger provided and observe charging instructions
Disposal
Always dispose of batteries responsibly according to local government guidelines.
49
20. TROUBLESHOOTING
If your
Flexistim
is not working properly please check the following:
Problem:
No display/won’t turn on:
BATTERY:
i) Is it fitted?
ii) Is it charged?
Controls don’t work
i) Press Ch1 or Ch2 ▼button to unlock
the keypad
ii) No showing. Ensure battery is charged.
No impulse output from electrodes
A circuit is not being made.
i) Have you applied both electrode pads (per
lead wire) to ensure a complete circuit?
ii) Are the lead wires properly connected at both
ends?
iii) Is the lead damaged? (Try using the other
lead - if this works, then the original lead is faulty)
Warning triangle flashing,
You are in EMS Manual Programme 10 or 11. When
cannot increase intensity.
using ALTERNATING, ensure that the “Ramp Up
Time”, “Work Time”, and “Ramp Down Time” equals
“Rest Time”. These will appear on your screen if this
formula is not followed in EMS manual mode.
In EMS Manual programmes10 or 11 a Warning triangle
will be displayed if the Rest period is less than Rest
Time=Work Time X (WorkHz-16.66)/16.66 seconds.
In IF MODE when intensity reaches 40mA, the
warning triangle flashes and intensity cannot
be increased.
Check electrode pad position. If you are certain that
the pads are positioned safely, press the SKIP key to
override. The triangle will stop flashing and intensity
can be increased to 60mA.
If the above review has failed to resolve your problem, call TensCare or your local dealer
for advice.
50