17
Treatment Timer
When the Treatment Timer has been set, it begins to count down in minutes and the
time remaining is displayed on the LCD. When it reaches zero, the device automatically
shuts off.
Memory
There is a usage memory for each Mode.
Pressing the
MODE
and
CH1
▼ button together for 3 seconds enters Memory.
The Usage time will be accumulatively recorded when the output level is above zero. The
accumulative treatment time in minutes is displayed.
Press the same keys again to return back to the previous normal display.
Pressing
CH2
▼ key and
MODE
key together for 3 seconds will clear the treatment
time to zero.
Power Supply
The
Flexistim
may be operated from the re-chargeable battery, or
directly from the power adaptor. When the adaptor is plugged in to
the
Flexistim
, the internal battery is automatically disconnected.
The battery cannot be charged while in the unit, only in the charging
cradle supplied.
The rechargeable battery will give about 1 hours use at 50% intensity in
IFT Mode, and much longer in other Modes.
If you need to make more than one treatment you may either:
a) Purchase and recharge additional batteries
b) Connect to mains power using the mains adaptor
18
9. CONTENTS OF THE PACK
Your
Flexistim
pack should contain the following:
1 ×
Flexistim
Unit
2 × Leads
4 × Self-Adhesive Electrodes with Connectors (Size: 50mm×50mm)
1 × 3.7V Rechargeable Lithium Battery (BL-6F)
1 × AC Power Cord
1 × Battery Charging Cradle
OTHER FEATURES
1. The LCD is backlit. To save energy the back light will switch off if the keypad is not
used for 30 secs. Pressing any key will re-activate it.
2. When the unit is turned on, if you do not press any of the keys, or intensity is set to
zero, for > 5 mins it will automatically shut off.
3. When you turn the unit on, it will automatically enter the mode that you used last.
4. When you change Programme, the output level will reset to zero immediately.
5. When the battery is low, the battery icon will flash indicating that the batteries should
be recharged.
WARNING: The power adaptor supplied has special medical grade isolation.
Use of any mains adaptor other than the one supplied with the device could
compromise electrical safety.