8
Replacing the Fuses.
Warning:
To avoid electrical shock,
personal injury or damage to the meter,
use specified fuses ONLY in accordance
with the following procedure.
1. Turn the meter power off and
remove all connections from the
terminals.
2. Remove the screw from the battery
compartment, and separate the
battery compartment from the case
bottom.
3. Remove the screws from the case
bottom, and separate the case top
from the case bottom.
4. Remove the fuse by gently prying
one end loose, then take out the
fuse from its holder.
5. Install ONLY replacement fuses
with the identical type and specification as follows and make sure the fuse is fixed
firmly in it’s holder. Fuse 1: 0.315A, 250V, fast type fuse, 5x20 mm.
6. Replace the battery compartment and the case top, and reinstall the screw.
7. Replace the case bottom and case top, and reinstall the screws.
Note:
Fuse replacement is seldom required. Blown fuses are typically a result of
improper use.
Screw