11
Inductance Test
Range
Resolution
Accuracy
Testing Frequency/Current
2mH
0.001mH
±(2%+8)
1kHz/150µA
20mH
0.01mH
200mH
0.1mH
2H
0.001H
±(5%+5)
20H
0.01H
±(5%+15)
100Hz/15µA
Measure of Inductance: 1H=103mH = 106μH.
Overload Protection: 0.315A, 250V, fast type fuse, 5x20 mm
Inductance Test
Range
Resolution
Accuracy
Testing Frequency/Voltage
2.000nF
0.001nF
±(1%+5)
1kHz/150mV
20.00nF
0.01nF
200.0nF
0.1nF
2.000µF
0.001µF
±(4%+5)
20.00µF
0.01µF
100Hz/15mV
200.0µF
0.1µF
600µF
0.001mF
Reference Only
100Hz/1.5mV
Overload Protection: 0.315A, 250V, fast type fuse, 5x20 mm
Measure of Capacitance: 1F=103mF = 106μF= 109nF = 1012pF
Discharge all high-voltage capacitors before testing capacitance.
CLEANING
•
Periodically wipe the case with damp cloth and mild detergent. Do not use
abrasives or solvents for cleaning.
•
Clean the measurement probe tips occasionally, as dirt on the probes can affect
reading accuracy.