62
Remarks:
z
z
Display:
a) True rms are valid from 10 of range to 100 of range
b) AC crest factor can be up to 3.0.
c) A residual reading of 80 digits with test leads shorted, will not affect stated accuracy.
d) When frequency is lower than 100kHz, the accuracy guarantee range 10 -100 .
e) When making AC+DC measurment, the accuray need to add (1 + 35 digits) of reading based on the above table.
At 10A range:
a) When the measured current is 5A, continuous measurement is allowed.
b) When the measured current is between 5A-10A, continuous measurement
10 seconds and interval more
than 15 minutes.
Range
Resolution
200µA
2000µA
20mA
200mA
10A
Bandwidth
Accuracy
72-7730A 72-7732A
Overload Protection
72-7730A 72-7732A
400µA
4000µA
40mA
400mA
10A
D. AC Current (AC+DC measurement is available)
0.01µA
0.1µA
0.001mA
0.01mA
0.001A
45Hz~1kHz
1kHz~10kHz
45Hz~1kHz
1kHz~10kHz
(0.8 +15)
(1.5 +40)
(2 +20)
(6 +40)
0.5A, 1000V, fast type
fuse,
10.3X38mm
10A, 1000V, fast type
fuse,
6.3X32mm
Model 72-7730A/72-7732A: OPERATING MANUAL
(0.7 +15)
(1 +40)
(1.5 +20)
(5 +40)