38
3.
The
BLUE
button cycles among resistance, continuity,
and diode.
For forward voltage drop readings on any semiconductor
component, place the red test lead on the componentís
anode and place the black test lead on the componentís
cathode. The red test lead polarity is + while the black
test lead polarity is - .
The measured value shows on the display.
Note
In a circuit, a good diode should still produce a
forward voltage drop reading of 0.5V to 0.8V; however,
the reverse voltage drop reading can vary depending
on the resistance of other pathways between the
probe tips.
Connect the test leads to the proper terminals as
said above to avoid error display.
The LCD will display
OL
indicating either open circuit
or wrong polarity connection.
The unit of diode is volt (V), displaying the positive-
connection voltage-drop value.
O p e n c i r c u i t v o l t a g e a p p r o x i m a t e 2 . 8 V.
l
l
l
l
l
l
When diode testing has been completed, disconnect
the connection between the test leads and the circuit
under test and remove test leads from the input
terminals of the Meter.
Model 72-7730A/72-7732A: OPERATING MANUAL
Содержание 72-7730A