Tenma 72-6854 Instruction manual  

The output will also be tripped off if an attempt is made to draw power from the sense wires. 

When the output is tripped the OUTPUT lamp will still be ON but the displays will show ‘OP trip’ 
and the UNREG lamp will also light. Turn the output off; the trip message should be replaced with 
the normal preset V and I readings. When the cause of the trip has been removed the output can 
be switched on again. 

Even with the output off the load is still connected to the power supply output stage. Do not apply 
external voltages in excess of 50V to the power supply terminals or damage may result. 

The output is protected from reverse voltages by a diode; the continuous reverse current must not 
exceed 3 Amps, although transients can be much higher. 

Maintenance 

For service repair and calibration please contact your local Tenma distributor or go to 
www.tenma.com 

Fuse 

The correct fuse type is: 

10 Amp 250V HBC time-lag, 5 x 20mm. 

Make sure that only fuses of the required rated current and specified type are used for 
replacement. The use of makeshift fuses and the short-circuiting of fuse-holders is prohibited. 

To replace the fuse, first disconnect the instrument from the AC supply. Remove the 6 cover 
securing screws and lift off the cover. Replace the fuse with one of the correct type and refit the 
cover. 

Note that the main function of the fuse is to make the instrument safe and limit damage in the 
event of failure of one of the switching devices. If a fuse fails it is therefore very likely that the 
replacement will also blow, because the supply has developed a fault; in such circumstances the 
instrument will need to be returned to the manufacturer for service.  

Cleaning 

If the PSU requires cleaning use a cloth that is only lightly dampened with water or a mild 
detergent. Polish the display window with a soft dry cloth. 

WARNING! TO AVOID ELECTRIC SHOCK, OR DAMAGE TO THE PSU, NEVER ALLOW 
WATER TO GET INSIDE THE CASE. TO AVOID DAMAGE TO THE CASE OR DISPLAY 
WINDOW NEVER CLEAN WITH SOLVENTS. 

 8 

www.tenma.com 

Содержание 72-6854

Страница 1: ...DUAL 35V 10A POWER FLEX POWER SUPPLY 72 6854 INSTRUCTION MANUAL ...

Страница 2: ...ded stays within the power envelope see graph Outside of this envelope the output becomes unregulated Output Switch Electronic Preset voltage and current displayed when off Output Terminals 4mm terminals on 19mm 0 75 pitch 30A max Sensing Switchable between local and remote Spring loaded push terminals for remote connection Output Impedance Typically 5mΩ in constant voltage mode Typically 5kΩ in c...

Страница 3: ...de lamp Constant current mode lamp Unregulated power limit lamp Trip message on display METER SPECIFICATIONS Meter Types Dual 4 digit meters with 12 5mm 0 5 LEDs Reading rate 4 Hz Meter Resolutions 10mV 10mA Meter Accuracies Voltage 0 2 of reading 1 digit Current 0 5 of reading 1 digit GENERAL AC Input 110V 120V AC or 220V 240V 10 50 60Hz Installation Category II Power Consumption 600VA max Operat...

Страница 4: ... interruption is prohibited The protective action must not be negated by the use of an extension cord without a protective conductor When the instrument is connected to its supply terminals may be live and opening the covers or removal of parts except those to which access can be gained by hand is likely to expose live parts The apparatus shall be disconnected from all voltage sources before it is...

Страница 5: ...tested Performance A According to EN61326 the definitions of performance criteria are Performance criterion A During test normal performance within the specification limits Performance criterion B During test temporary degradation or loss of function or performance which is self recovering Performance criterion C During test temporary degradation or loss of function or performance which requires o...

Страница 6: ...llows The wire which is coloured green and yellow must be connected to the terminal in the plug which is marked by the letter E or by the safety earth symbol shown above or coloured green or green and yellow The wire which is coloured blue must be connected to the terminal which is marked with the letter N or coloured black The wire which is coloured brown must be connected to the terminal which i...

Страница 7: ...lower meter now shows the actual load current Constant Voltage The output voltage is adjusted using the coarse and fine VOLTAGE control the CURRENT control sets the maximum current that can be supplied The CV lamp lights to show constant voltage mode Constant Current If the load resistance is low enough such that at the output voltage set a current greater than the current limit setting would flow...

Страница 8: ...lso be beneficial The voltage drop in each output lead must not exceed 0 5 Volts Switch the LOCAL REMOTE switch back to LOCAL when remote sensing is not in use Series or Parallel connection with other units The outputs of the power supply are fully floating and may be used in series with other power supply units to generate high DC voltages up to 300V DC WARNING Such voltages are exceedingly hazar...

Страница 9: ... 10 Amp 250V HBC time lag 5 x 20mm Make sure that only fuses of the required rated current and specified type are used for replacement The use of makeshift fuses and the short circuiting of fuse holders is prohibited To replace the fuse first disconnect the instrument from the AC supply Remove the 6 cover securing screws and lift off the cover Replace the fuse with one of the correct type and refi...

Страница 10: ...TEST EQUIPMENT 405 Pioneer Blvd Springboro Ohio 45066 Book Part No 48511 0420 Issue 5 ...

Отзывы: