EMC 

This instrument has been designed to meet the requirements of the EMC Directive 

2004

/

108

/EC. 

Compliance was demonstrated by meeting the test limits of the following standards: 

Emissions 

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use.  Test limits used were: 

a) Radiated: 

 

Class 

b) Conducted: 

Class 

c)  Harmonics: 

EN61000-3-2 (2000) Class A; the instrument is Class A by product  

 

  category. 

Immunity 

EN61326 (1998) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use. 

Test methods, limits and performance achieved were: 

a)  EN61000-4-2 (1995) Electrostatic Discharge : 4kV air, 4kV contact, Performance A. 

b)  EN61000-4-3 (1997) Electromagnetic Field, 3V/m, 80% AM at 1kHz, Performance A. 

c)  EN61000-4-11 (1994) Voltage Interrupt, 1 cycle, 100%, Performance B. 

d)  EN61000-4-4 (1995) Fast Transient, 1kV peak (AC line), 0.5kV peak (DC Outputs), 

Performance A. 

e)  EN61000-4-5 (1995) Surge, 0.5kV (line to line), 1kV (line to ground), Performance A. 

f) 

EN61000-4-6 (1996) Conducted RF, 3V, 80% AM at 1kHz (AC line only; DC Output 
connections <3m not tested), Performance A. 

According to EN61326 the definitions of performance criteria are: 

Performance criterion A:

  ‘During test normal performance within the specification limits.’ 

Performance criterion B:  

‘During test, temporary degradation, or loss of function or 

performance which is self-recovering’. 

Performance criterion C:

  ‘During test, temporary degradation, or loss of function or 

performance which requires operator intervention or system reset occurs.’ 

Where Performance B is stated it is because DC Output regulation may deviate beyond 
Specification limits under the test conditions.  However, the possible deviations are still small and 
unlikely to be a problem in practice. 

Note that if operation in a high RF field is unavoidable it is good practice to connect the PSU to 
the target system using screened leads which have been passed (together) through an absorbing 
ferrite sleeve fitted close to the PSU terminals. 

Cautions 

To ensure continued compliance with the EMC directive observe the following precautions: 

a)  after opening the case for any reason ensure that all signal and ground connections are 

remade correctly and that case screws are correctly refitted and tightened. 

b)  In the event of part replacement becoming necessary, only use components of an identical 

type, see the Service Manual. 

 

 4 

www.tenma.com 

Содержание 72-6854

Страница 1: ...DUAL 35V 10A POWER FLEX POWER SUPPLY 72 6854 INSTRUCTION MANUAL ...

Страница 2: ...ded stays within the power envelope see graph Outside of this envelope the output becomes unregulated Output Switch Electronic Preset voltage and current displayed when off Output Terminals 4mm terminals on 19mm 0 75 pitch 30A max Sensing Switchable between local and remote Spring loaded push terminals for remote connection Output Impedance Typically 5mΩ in constant voltage mode Typically 5kΩ in c...

Страница 3: ...de lamp Constant current mode lamp Unregulated power limit lamp Trip message on display METER SPECIFICATIONS Meter Types Dual 4 digit meters with 12 5mm 0 5 LEDs Reading rate 4 Hz Meter Resolutions 10mV 10mA Meter Accuracies Voltage 0 2 of reading 1 digit Current 0 5 of reading 1 digit GENERAL AC Input 110V 120V AC or 220V 240V 10 50 60Hz Installation Category II Power Consumption 600VA max Operat...

Страница 4: ... interruption is prohibited The protective action must not be negated by the use of an extension cord without a protective conductor When the instrument is connected to its supply terminals may be live and opening the covers or removal of parts except those to which access can be gained by hand is likely to expose live parts The apparatus shall be disconnected from all voltage sources before it is...

Страница 5: ...tested Performance A According to EN61326 the definitions of performance criteria are Performance criterion A During test normal performance within the specification limits Performance criterion B During test temporary degradation or loss of function or performance which is self recovering Performance criterion C During test temporary degradation or loss of function or performance which requires o...

Страница 6: ...llows The wire which is coloured green and yellow must be connected to the terminal in the plug which is marked by the letter E or by the safety earth symbol shown above or coloured green or green and yellow The wire which is coloured blue must be connected to the terminal which is marked with the letter N or coloured black The wire which is coloured brown must be connected to the terminal which i...

Страница 7: ...lower meter now shows the actual load current Constant Voltage The output voltage is adjusted using the coarse and fine VOLTAGE control the CURRENT control sets the maximum current that can be supplied The CV lamp lights to show constant voltage mode Constant Current If the load resistance is low enough such that at the output voltage set a current greater than the current limit setting would flow...

Страница 8: ...lso be beneficial The voltage drop in each output lead must not exceed 0 5 Volts Switch the LOCAL REMOTE switch back to LOCAL when remote sensing is not in use Series or Parallel connection with other units The outputs of the power supply are fully floating and may be used in series with other power supply units to generate high DC voltages up to 300V DC WARNING Such voltages are exceedingly hazar...

Страница 9: ... 10 Amp 250V HBC time lag 5 x 20mm Make sure that only fuses of the required rated current and specified type are used for replacement The use of makeshift fuses and the short circuiting of fuse holders is prohibited To replace the fuse first disconnect the instrument from the AC supply Remove the 6 cover securing screws and lift off the cover Replace the fuse with one of the correct type and refi...

Страница 10: ...TEST EQUIPMENT 405 Pioneer Blvd Springboro Ohio 45066 Book Part No 48511 0420 Issue 5 ...

Отзывы: