The reference profile section is marked with vertical  lines  called the average
reference cursors.  These cursors are drawn with large dashes and connected by
bars at the top and bottom of the screen to distinguish them from the other
cursors.

1.

 

Position the measurement cursors  on  the  profile section to be compared to
the reference.

The field labeled "Avg" will show the average profile  height between the
measurement cursors.  The field labeled "dAv" will now show  the  difference
between  this  value  and  the  reference reading.

When the measurement cursors are moved, the  average height and average
difference will be updated.

2.

 

To exit Average Difference mode,  press  CUR  then ABT.   The vertical
height field will now display the difference between the  cursor  heights.   See
the  hard  copy  manual  for  a figure showing the Average Difference Mode
display.

        4.6.  Printing

There are three available printing modes: "Screen"  in which the entire contents
of the screen, including the plot of the profile, is printed; "Summary Data"
prints out  only  the  data  shown  on  the  left side of the screen; "Data Send"
sends  all  data  to  the  RS-232C

                serial output instead of the printer.

Screen and Summary Data print modes are used  to  make print-outs  on the
internal printer or on the optional remote printer.  Data Send  uses  the  RS-
232C  serial output  to send the data to an external device such as

               an 80-column serial printer or a host computer.

Your choice of printer mode remains  in  effect until changed.   A  printout  or
data send can be initiated whenever the results of a scan are  are  displayed  on

                 the screen.

1.

 

Initiate "Enter Mode" by hitting the  ENT  key  if you have not already done
so.

2.  Press the white print button to change  the  print mode.    The  selection  will
     be  highlighted  in reverse video.

3.

 

With the desired mode  selected,  press  ENT  to leave  the Scan Menu, or
press START/STOP to leave the Scan Menu and begin a scan.

Содержание AlphaStep 200

Страница 1: ...e scan as well as scan profiles and a summary of scan data The system has three operating modes manual for single scans semiautomatic for multi point scans and automatic for repeated multi point scans The system allows the operator to set and store if desired a variety of scan parameters such as scan direction scan length sampling rate and vertical units kA vs Mm Measurement cursors are used to le...

Страница 2: ...ed a plot of the scan is displayed along with pertinent data and information on the scan characteristics The cursors displayed on the plot are positioned by the operator to level the plot and or obtain the desired data The figure in the hard copy manual shows what the display looks like The following describes the data display as read from top to bottom ID Optional identifier for sample assigned b...

Страница 3: ...ple density specifications scan time scan direction stylus force and the vertical units 2 4 Keyboard There is a figure showing the keyboard in the hard copy manual A brief description of key functions follows For further information on using the keyboard see the detailed operating procedures in Section 3 ENT Enter Pressing this button will initiate terminate Enter Mode in which the operator can ch...

Страница 4: ... change as either the or down arrow key is pressed PLOT Used to zoom in on a section of a profile between the measurement cursors or to level a profile CUR This key and its associated and keys allow the operator to set scan direction and place the measurement cursors LEVEL Used to level the baseline of a profile TABLE The and down arrow keys will move the sample chuck up and down respectively PRIN...

Страница 5: ...een You should see your sample come into view and the stylus lower and come to rest on the surface of your substrate If you continue to press TABLE the stylus will retract Thus the TABLE button toggles the stylus between its fully retracted and lowered position Always retract the stylus when moving the sample 6 Use the zoom knob to adjust the magnification of your sample Use the stage arrow keys t...

Страница 6: ...data summary on the left side of the screen The CUR button selects the cursor to be moved You can see the active cursor indicator appear over each cursor in turn and then over both as you repeatedly press this button When both cursors are active they will remain a fixed distance apart as they move simultaneously The arrow buttons will move the cursor s in the direction indicated Holding down these...

Страница 7: ...lly located at the ends other scan If one cursor is on a surface feature higher or lower than the other cursor the substrate will appear tilted when the program makes both cursor heights equal The condition described above can be corrected by moving the leveling cursors to places on the profile where the heights are equal Ideally these should both be on the substrate near the ends of the scan alth...

Страница 8: ...e screen showing the entire profile press ABT This undoes all zooming but leaves the cursors at their most recent positions on the profile Pressing CUR then ABT undoes all zooming and returns the cursors to the ends of the scan 4 5 Average Difference Mode In Average Difference mode the system computes the difference between average height measurements When this mode is entered the currently displa...

Страница 9: ...y 4 6 Printing There are three available printing modes Screen in which the entire contents of the screen including the plot of the profile is printed Summary Data prints out only the data shown on the left side of the screen Data Send sends all data to the RS 232C serial output instead of the printer Screen and Summary Data print modes are used to make print outs on the internal printer or on the...

Страница 10: ...that the paper unwinds from underneath as it leaves the spool This is thermal paper if you do not have the correct side facing the printer nothing will be printed on the paper 5 Insert the paper underneath the rubber roller from behind This is easier if the end of the paper is cut so that it tapers to a point 6 The paper will emerge at the top of the rubber roller aiming back Pull out several inch...

Отзывы: