8
7)
After connecting strips cover
ceramic board with insulation
wool for prevention heat loses.
8)
Insert the Controlling Sensor in chamber till it
doesn't touch the ceramic tube. Controlling Sensor is
very fragile please insert it very carefully.Sensor
sholud be touch the ceramic tube otherwise deviation
will occur.
Insert controlling sensor here
Insert sensor in the hole carefully
tight bolt with screw driver
Note :
remember that
after inserting B type
sensor screw should
not be very tight and
sensor not insert before
insertion of heating
e l e m e n t o t h e r w i s e
sensor will be damage.
9)
After above all steps complete. The inner chamber
is Ready for installation in outer cabinate.
The inner chamber insert carefully from front side and
after this all four nut bolt tight shown as below.
fix the ss chamber into
outer cabinate
nut bolt tight
12)
After this the transformer lead connect (through
nut and bolt with the help of spanner) in the open strip
from back side as per shown in Figure.
cable connect
with right open
strip in back side
cable connect
with left open
strip in back
side
13)
Screw tight the cable with the strips . As shown in
figure.
10)
Check the ss chamber proper fit in the outer
chamber.
14)
After all Connection (Not Loose also not too much
given force to terminals ) the Back View of furnace is
Look As below.
Note :
Once you heated 3-4 times then recheck all
heater connection of CALsys 1700 (there are no any
loose connection) to avoid any drops of voltage.
Re-fix the front panel and back with help of bolts
11)
Connect B type sensor M & S terminal with safety
controller as shown in the Figure.
Sensor
M
Terminal
Sensor
S
Terminal