will   switch   the   CALsys   1500BB   ON   or   OFF   to   raise   or   lower   the   temperature.   The

displayed temperature will gradually change until it reaches the setpoint temperature.

The temperature bath requires 15 to 20 minutes to reach the set-point depending on the

span. Another 10 to 15 minutes is required to stabilize the temperature bath within ±

1.0ºC of the set-point.

5.3 Operating Instruction:- 

1.

after the completly installation.

2.

Connect the ‘CALsys 1500BB’ to a suitable power supply. ON the Mains switch located at back

side with power entry.

3.

set the desire temperature value in PID by using UP & Down key.

4.

Aim the reference (Master) standard infrared temperature sensor to the target area.

5.

PV   (Present   value)   display   in   controller   will   gradually   rise   until   it   reaches   the   set   point

temperature. The  

controller takes some times to reach the set-point depending on the span.

Furnace is stable when PV is equal to SV (Set Value).

6.

Master sensor take some times to reach the setpoint temperature and stable at temperature

near about controller  set temperature.

7.

When   temperature   of   the   master   and  UUC   (Unit   under   calibration)   are  stable   record   the

readings of master sensor.

8.

Compare the UUC reading with the master’s reading & find out the error by comparison method.

9.

Reset the controller and / or repeat the calibration for another calibration point or for another

sensor.

10.

When the calibration is complete, reset the controller to 0°C & wait until the unit has cooled to

below 100°C, before moving the ‘CALsys 1500BB to new location the ‘CALsys 1500BB’ must be

cooled below 100°C before it can be put back into its carrying case.

11.

NOTE: Always use a reference IR thermometer for comparison calibration method.

Note:

All other controller parameters are company set and locked. It is 

recommended not to change them.

When the source is operated at any temperature above ambient, 

the front face and plate become hot.

UP

Key

Down

Key

Содержание CALsys 1500 BB

Страница 1: ...CALsys 1500 BB CALsys 1500 BB Black Body Temperature Calibrator USER MANUAL...

Страница 2: ...on 3 2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 T...

Страница 3: ...through defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplie...

Страница 4: ...the flexible mains cable which is colored green yellow MUST be connected to a suitable earth Warning Internal mains voltage hazard Do not remove the panels There are no user serviceable parts inside C...

Страница 5: ...230 V AC Power Maximum Supply Frequency 50 60 Hz Temperature range 500 to 1500 C Resolution 1 0 C Stability 1 0 C Controlling sensor R type T C Time to reach max Temperature 2 5 hours Operating Temper...

Страница 6: ...other pro ducts DANGER Power is present even if the front panel circuit breaker is in the OFF 0 position 4 2 Outline Description Calsys 1500BB is a transportable unit designed for use on any reasonab...

Страница 7: ...ed same sr no On heater holder and chamber NOTE HEATERS INSERT VERY CAREFULLY After Inserting its tight by screwdriver Open back panel sheet of CALsys 1500 and remove all Packing sheet from inside Ins...

Страница 8: ...ven picture Open holding ceramic clamp as shown in figure Open the front plate of CALsys 1500BB and remove all packing sheets from inside Fix the 6 heater into the chamber and Connect all the aluminum...

Страница 9: ...ensure that the tail of cavity fix to the groove on the back panel And make sure cavity properly insert at back panel hole as shown in image After insert the cavity tight the holding cap of SS chambe...

Страница 10: ...5 1 Power Plug the black body power cord into a mains outlet of the proper voltage frequency and current capability Typically this will be 230 VAC 10 50 60 Hz Turn on the black body using Mains switc...

Страница 11: ...n the span Furnace is stable when PV is equal to SV Set Value 6 Master sensor take some times to reach the setpoint temperature and stable at temperature near about controller set temperature 7 When t...

Страница 12: ...the set point to the required value The lower display changes to indicate the new set point 6 3 Monitoring the Controller Status A row indicate the controllers status as follows OP1 Heat Output REM Th...

Страница 13: ...s recommended Cleaning may be accomplished by the use of a small dry paint brush 7 2 Trouble shooting 1 Unit fails to operate Check fuse if it is tripped switch is ON If not power ON of calsys 1500 BB...

Страница 14: ...f software 1 2 Parameter in mainscreen 1 2 1 Communication Communication between the furnace and the software is implemented via a RS 232 cable connected between the furnace and the PC serial port Thi...

Страница 15: ...ging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVI Value Read th...

Страница 16: ...el to protect it from humidity Warranty TEMPSENS CALsys 1500BB instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are not cover...

Страница 17: ......

Отзывы: