Tempsens CALsys 1500 BB Скачать руководство пользователя страница 1

CALsys 1500 BB

CALsys 1500 BB

 Black Body Temperature Calibrator

 USER MANUAL

Содержание CALsys 1500 BB

Страница 1: ...CALsys 1500 BB CALsys 1500 BB Black Body Temperature Calibrator USER MANUAL...

Страница 2: ...on 3 2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 T...

Страница 3: ...through defective material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplie...

Страница 4: ...the flexible mains cable which is colored green yellow MUST be connected to a suitable earth Warning Internal mains voltage hazard Do not remove the panels There are no user serviceable parts inside C...

Страница 5: ...230 V AC Power Maximum Supply Frequency 50 60 Hz Temperature range 500 to 1500 C Resolution 1 0 C Stability 1 0 C Controlling sensor R type T C Time to reach max Temperature 2 5 hours Operating Temper...

Страница 6: ...other pro ducts DANGER Power is present even if the front panel circuit breaker is in the OFF 0 position 4 2 Outline Description Calsys 1500BB is a transportable unit designed for use on any reasonab...

Страница 7: ...ed same sr no On heater holder and chamber NOTE HEATERS INSERT VERY CAREFULLY After Inserting its tight by screwdriver Open back panel sheet of CALsys 1500 and remove all Packing sheet from inside Ins...

Страница 8: ...ven picture Open holding ceramic clamp as shown in figure Open the front plate of CALsys 1500BB and remove all packing sheets from inside Fix the 6 heater into the chamber and Connect all the aluminum...

Страница 9: ...ensure that the tail of cavity fix to the groove on the back panel And make sure cavity properly insert at back panel hole as shown in image After insert the cavity tight the holding cap of SS chambe...

Страница 10: ...5 1 Power Plug the black body power cord into a mains outlet of the proper voltage frequency and current capability Typically this will be 230 VAC 10 50 60 Hz Turn on the black body using Mains switc...

Страница 11: ...n the span Furnace is stable when PV is equal to SV Set Value 6 Master sensor take some times to reach the setpoint temperature and stable at temperature near about controller set temperature 7 When t...

Страница 12: ...the set point to the required value The lower display changes to indicate the new set point 6 3 Monitoring the Controller Status A row indicate the controllers status as follows OP1 Heat Output REM Th...

Страница 13: ...s recommended Cleaning may be accomplished by the use of a small dry paint brush 7 2 Trouble shooting 1 Unit fails to operate Check fuse if it is tripped switch is ON If not power ON of calsys 1500 BB...

Страница 14: ...f software 1 2 Parameter in mainscreen 1 2 1 Communication Communication between the furnace and the software is implemented via a RS 232 cable connected between the furnace and the PC serial port Thi...

Страница 15: ...ging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVI Value Read th...

Страница 16: ...el to protect it from humidity Warranty TEMPSENS CALsys 1500BB instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are not cover...