4. Overview

4.1 Introduction:

The ‘

CALsys 1200’ 

has been designed to provide stable and accurate temperature to enable

professionals   to   calibrate   Temperature   Sensing   Devices   by   comparison   method.   The

CALsys 1200’

  model has been designed to  be rugged and easily maintained. This model

provides   an   isothermal   enclosure   (Metal   block)   in   which   the   thermocouple/RTD   can   be
calibrated   against   the   temperature   of   the   calibrator.   For   traceable   calibration   a   master
calibration sensor should be placed into the metal block alongside the unit under calibration.
The   method   is   widely   accepted   because   the   calibrator   provides   very   stable   temperature
nearing   to   its   controlled   point,   the   Master   Thermocouple   /   RTD,   which   is   calibrated   by
independent   Laboratory   traceable   to   National   standards,   compares   the   sensor   under
calibration.
The ‘

CALsys

’ models are part of wide range of portable calibrators designed and made by us.

Please contact us in case you required more information about our other products.

4.2 Outline Description:

‘The calibrator controller uses a precision R type thermocouple as a controlling sensor and
controls the well temperature with Kanthal spiral heater. To obtain and maintain a required
temperature the controller varies the power to the heater via solid-state relay. There is one
electricity driven fan which is situated under the heating chamber for cooling the heater. 

The

CALsys   1200   dry   block   calibrator   was   designed   for   portability,   moderate   cost   and   ease   of
operation.  With  proper   use  the  instrument  should  provide  continued  accurate   calibration  of
temperature sensors and devices. The user should be familiar with the safety guidelines and
operating procedures of the calibrator as described in the User’s Manual.

5. Operating Instruction

(1)

PID Controller

(2)

ON/OFF switch of heater

(3)

Power cable

(4)

RS-232 output

Содержание CALsys 1200

Страница 1: ...CALsys 1200 CALsys 1200 Dry block Temperature Calibrator USER MANUAL TEMPSENS INSTRUMENT I PVT LTD UNIT II...

Страница 2: ...2 Circuit diagram 4 Overview 4 4 1 Introduction 4 2 Outline description 5 Operating Instruction 5 5 1 Power 5 2 Setting the temperature 5 3 Operating Instruction 6 Operation of Controller 6 6 1 The T...

Страница 3: ...ctive material or workmanship provided the failure is not the result of misuse In the event of failure covered by this warranty the instrument must be returned carriage paid to the supplier for examin...

Страница 4: ...for repair Do not use the apparatus outside its recommended range i e 250 to 1200 C Ensure inflammable materials are kept away from hot parts of the apparatus to prevent fire risk Before connecting to...

Страница 5: ...ge 250 to 1200 C Resolution 0 1 C Stability 0 5 C Controlling sensor R type T C Time to reach max Temperature 90 Min Operating Temperature 20 to 45 C Controller Specifications Eurotherm Stabilization...

Страница 6: ...The CALsys models are part of wide range of portable calibrators designed and made by us Please contact us in case you required more information about our other products 4 2 Outline Description The c...

Страница 7: ...ze the temperature bath within 0 5 C of the set point 5 3 Operating Instruction 1 Insert the inconel block into well of furnace 2 Connect the CALsys 1200 to a suitable power supply Switch ON the Dry b...

Страница 8: ...point to the required value The lower display changes to indicate the new set point 6 3 Monitoring the Controller Status A row indicate the controllers status as follows OP1 Heat Output OP2 Cool Outpu...

Страница 9: ...recommended Cleaning may be accomplished by the use of a small dry paint brush 7 2 Trouble shooting 1 Unit fails to operate Check fuse if it is tripped switch is ON If not power ON of calsys 1200 bat...

Страница 10: ...f software 1 2 Parameter in mainscreen 1 2 1 Communication Communication between the furnace and the software is implemented via a RS 232 cable connected between the furnace and the PC serial port Thi...

Страница 11: ...ging up to 120 minutes than clickon start graph button After complete the task clickon save to file button Set Point C in which you can set temperature of furnace as your requirement PVI Value Read th...

Страница 12: ...th silicon gel to protect it from humidity Warranty TEMPSENS CALsys 1200 instrument have a warranty of one year from the invoice date This warranty covers manufacturing defects User induced faults are...

Страница 13: ...TEMPSENS INSTRUMENT I PVT LTD...

Отзывы: