AFA1000/1 MK2 Airflow Monitor
Installation & Operating Manual
5.1 Installation of standard components
The following section outlines the installation of the various components of the Airflow
monitor system.
As the size and format of individual Fume Cupboards varies considerably, specific
instructions are not possible, though the principles outlined below should remain valid
in all cases.
The AFA1000/1 monitor can be mounted on either side of the Fume Cupboard;
consideration should be made for the cable lengths when fitting the airflow sensor.
1.
Fit the AFA1000/1 airflow monitor to the Fume Cupboard using the cut-out details
provided with the unit --- see page 25. Ensure that the monitor fits into the cut out
before securing with the fixing screws.
Separate mounting boxes and adaptor plates are available for retro-fitting to older
Fume Cupboards.
2.
Fit the Plug in supply to the top of the Fume Cupboard ensuring that the location
allows the 5M low voltage cable to reach the AFA1000/1 airflow monitor terminals.
3.
Fit the airflow sensor to the Fume Cupboard using the cut out and installation details
provided --- see page 31. When possible ensure that the Airflow sensor is mounted
on the same side of the Fume Cupboard as the AFA1000/1 so that the standard 2M
sensor cable will reach the sensor and monitor.
4.
Connect the ‘telephone style’ airflow sensor plug-in cable to the sensor and the back
of the AFA1000/1 unit --- see typical connection diagram on page 25.
Longer Sensor Cables are available on request.
5 INSTALLATION
Содержание AFA1000/1 MK2
Страница 4: ...1 SAFETY...
Страница 9: ...2 OVERVIEW OF THE AIRFLOW MONITOR...
Страница 19: ...3 FUNCTIONS OPERATION...
Страница 24: ...4 COMPONENTS...
Страница 27: ...5 INSTALLATION...
Страница 32: ...6 CONFIGURATION CALIBRATION...
Страница 36: ...7 AUXILIARY FEATURES CONNECTIONS...
Страница 41: ...8 RS485 COMS OUTPUT...
Страница 48: ...9 WARRANTY...