AFA1000/1 MK2 Airflow Monitor
Installation & Operating Manual
External alarm
• When input configured as External alarm is activated
• Red LED on (Flashing) – (if configured)
• External Alarm toggles on /off with display -- (if configured)
• Audible alarm sounds – can be muted via Enter pushbutton
• External alarm relay operates (if configured)
Emergency
• When input configured as Emergency is activated
• Red LED on (Flashing) – (if configured)
• Emergency toggles on /off with display -- ( if configured )
• Audible alarm sounds – can be muted via Enter pushbutton
• Emergency alarm relay operates (if configured)
Sash High
• When input configured as Sash High is activated
• Amber LED on Solid (Flashing when alarm muted)
• Sash High toggles on/off with display if sash height > calibrated position
• Audible alarm sounds -- can be muted via Enter pushbutton
• Sash High relay operates (if configured)
• Alarm re-activates after time delay if repeat time enabled
High / Low
• When input configured as High/Low is activated
• Display Icon shows Up (High) or Down (Low) arrows
• High/Low relay operates (if configured)
This function is designed for two speed fan operation or two position damper
operation switched via a micro switch or proximity switch activated at a given position
on the sash.
Close Sash
• When the input configured as Close Sash is activated
• Red LED on
• Close Sash – toggles on/off with velocity display
• Audible alarm sounds (after pre-set time)
• Audible can be muted via Enter pushbutton -- this silences the alarm if configured.
Reset when Sash lowered to closed position and input de-activated or operator is
present
Содержание AFA1000/1 MK2
Страница 4: ...1 SAFETY...
Страница 9: ...2 OVERVIEW OF THE AIRFLOW MONITOR...
Страница 19: ...3 FUNCTIONS OPERATION...
Страница 24: ...4 COMPONENTS...
Страница 27: ...5 INSTALLATION...
Страница 32: ...6 CONFIGURATION CALIBRATION...
Страница 36: ...7 AUXILIARY FEATURES CONNECTIONS...
Страница 41: ...8 RS485 COMS OUTPUT...
Страница 48: ...9 WARRANTY...