3.2
Sensor and Cable Part Numbers
4
HIGH VACUUM VERSIONS AND OPERATION
3.2
Sensor and Cable Part Numbers
Table 10: MC300 Cable Part Numbers
Length
2A Cable
4A Cable
7B Cable
7E/F/FC Cable
Mini BA Cable
10 ft
2-9800-077 2-9820-010 2-9800-09
2-9841-010
2-9854-10
20 ft
2-9800-078 2-9820-020 2-9800-41
2-9841-020
2-9854-20
35 ft
2-9800-079 2-9820-035 2-9800-42
2-9841-035
2-9854-35
50 ft
2-9800-080 2-9820-050 2-9800-43
2-9841-050
2-9854-50
Table 11: Selected Sensor Part Numbers
Sensor and Flange Type
Part Number
2A 1/8" NPT Brass
2-2100-10
2A 1/8" NPT Stainless Steel
2-2126-001
2A NW16 Stainless Steel
2-2126-030
4A 1/8" NPT Brass
2-2119-001
4A 1/8" NPT Stainless Steel
2-2120-001
4A NW16 Stainless Steel
2-2120-030
7B 3/4" NPT Brass
2-2100-263
7B NW25 Stainless Steel
2-2100-272
7E NW25
2-2141-031
7F 2 3/4" Conflat
2-2145-052
7FC 2 3/4" Conflat
2-2146-053
Mini BA NW25
3451-8305-35
Mini BA 2 3/4" Conflat
3451-8305-25
4
High Vacuum Versions and Operation
High vacuum versions of the MC300 include automatic operation of the high vacuum sensors based
on the low vacuum sensor in Channel 1. Channel 2 is also a low vacuum sensor. The high vacuum
sensor is labeled "ION GAUGE" on the MC300 display and is referred to as Channel 3.
4.1
Cold Cathodes Configurations
For MC300 configurations that contain a cold cathode (either 7B or 7E/7F/7FC), the cold cathode
sensor will occupy Channel 3. When the low vacuum sensor pressure in Channel 1 is below 10 mTorr,
the high voltage to the cold cathode will be enabled and the sensor will produce a reading. If the
pressure is above the acceptable range for the cold cathode, it will display
OLd
to indicate this. When
the pressure in Channel 1 is above 10 mTorr, the cold cathode sensor will dispaly
OFF
and there will
be no high voltage to the sensor.
4.2
Mini Bayard-Alpert Configurations
For MC300s with Mini Bayard-Alpert configurations, there are four extra steps in the setup menu.
More informations is contained below.
Note for the BA that whenever the filament is turned off via the user or via over current protection,
the MC300 power must be cycled to turn on the mini BA sensor again.