![TELEVAC MC300 Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/televac/mc300/mc300_instruction-manual_1082535017.webp)
2.5
Analog Outputs
2
SETUP
Table 6: 7E/7F/7FC Analog Output
Pressure
(Torr)
Logarithmic
Output (V)
Linear by
Decade
Output (V)
1.00E-11
0
0.1
2.51E-11
0.4
0.251
6.31E-11
0.8
0.631
1.00E-10
1.0
1.1
2.51E-10
1.4
1.251
6.31E-10
1.8
1.631
1.00E-09
2.0
2.1
2.51E-09
2.4
2.251
3.98E-09
2.6
2.398
6.31E-09
2.8
2.631
1.00E-08
3.0
3.1
2.51E-08
3.4
3.251
3.98E-08
3.6
3.398
6.31E-08
3.8
3.631
1.00E-07
4.0
4.1
2.51E-07
4.4
4.251
3.98E-07
4.6
4.398
6.31E-07
4.8
4.631
1.00E-06
5.0
5.1
2.51E-06
5.4
5.251
3.98E-06
5.6
5.398
6.31E-06
5.8
5.631
1.00E-05
6.0
6.1
2.51E-05
6.4
6.251
3.98E-05
6.6
6.398
6.31E-05
6.8
6.631
1.00E-04
7.0
7.1
2.51E-04
7.4
7.251
3.98E-04
7.6
7.398
6.31E-04
7.8
7.631
1.00E-03
8.0
8.1
2.51E-03
8.4
8.251
3.98E-03
8.6
8.398
6.31E-03
8.8
8.631
1.00E-02
9.0
9.1
OL’D
>9.2
>9.160