![TELEVAC MC300 Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/televac/mc300/mc300_instruction-manual_1082535011.webp)
2.5
Analog Outputs
2
SETUP
Table 3: 2A Analog Output
Pressure
(Torr)
Traditional
Output (V)
Linear
Output (V)
-0.005
10.59
-0.05
-0.001
10.16
-0.01
0
10
0
0.001
9.84
0.01
0.005
9.41
0.05
0.01
8.92
0.1
0.02
8.03
0.2
0.03
7.23
0.3
0.04
6.63
0.4
0.05
6.13
0.5
0.06
5.69
0.6
0.07
5.31
0.7
0.08
4.95
0.8
0.09
4.63
0.9
0.1
4.37
1
0.15
3.37
1.5
0.2
2.89
2
0.3
2.23
3
0.4
1.88
4
0.5
1.67
5
0.6
1.58
6
0.7
1.5
7
0.8
1.42
8
0.9
1.34
9
1
1.26
10
2
1.02
10.2
5
0.92
10.2
10
0.9
10.2
20
0.88
10.2
CABLE
0
10.2
email: [email protected]
tel: +1 215 947 2500
web: frederickscompany.com
mc300_im rev H
10 of 23