High Voltage Shielded Cable
See Figure 3-C
Approximately 20 feet of RG-213/U high voltage coaxial cable is provided for connection
of the power supply output to the source transformer and is shipped connected to
emission control module.
Approximately 5 inches of outer covering should be removed, and the shield stripped back
so the insulated center conductor is exposed. The shield should then be connected to the
tank ground at the transformer. The same care should be taken in making these
connections as that taken in connecting earth ground to the tank and transformer.
The center conductor should be attached to the secondary center-tap on the top of Source
Filament transformer. The cable is supplied with appropriate connections.
You must have a HV cable for each source from the power supply.
In addition, you have been supplied two 3 foot lengths of special coaxial cable
(#6 AWG with 40KVDC dielectric insulation) to be used from the filament transformer
output to the vacuum tank high voltage feedthrus. If a different length is desired, please
contact the factory.
Inter-module Connections
See Figure 3-D or 3-E
Control to HV/Power
•
J4 of Controller to Power Supply “HV Control”
J4 is a 25-pin D-sub connector located on the rear panel of the HV module. The 25 wires
carry many signals both for commands and feedback information in analog form.
HV Power Supply Source Module to Sweep
•
Rectangle plug, 2-pin plus ground
This line carries single phase line voltage (220V) to the Sweeper Module.
Control Module
This section applies to all Control Modules, (Single or Multi-gun)
Interlock Connections - J1
The external interlocks are connected to the power supply through J1 (9 pin D-sub), which
is located on the rear of the control module. Each controller (HV Control/Source 1 and
Source 2/3) has an independent interlock circuit. See Figure 3-D or 3-E, Control Module
Connections for details. 22 gauge ribbon cable or round cable should be used for the
interlock installation that is described below.
•
Auxiliary Interlock 2 - Source 1, 2 and 3
INSTALLATION
3-5
Telemark TT-10/15 Power Supply
Содержание TT-10
Страница 9: ...GENERAL DESCRIPTION 1 3 Telemark TT 10 15 Power Supplies Figure 1 B Cabinet General View...
Страница 10: ...1 4 GENERAL DESCRIPTION Telemark TT 10 15 Power Supplies...
Страница 18: ...3 6 INSTALLATION Telemark TT 10 15 Power Supply Figure 3 D Source Inter Module Connections...
Страница 21: ...INSTALLATION 3 9 Telemark TT 10 15 Power Supply Figure 3 E Control Module Connections...
Страница 34: ...4 10 CONTROLS AND INDICATORS Telemark TT 10 15 Power Supplies...