Telemark TT Cheetah Series Power Supply Instruction Manual
16
Figure 4-H HVPS Rear Panel Connectors (TT-6 shown)
Interconnection Procedure
Figure 4-I Controller Rear Panel Connectors
1. Turn OFF all front panel circuit breakers.
2. Connect the ground stud on the rear panel of the HVPS chassis to a common ground on
the vacuum system. Use a 0.5 inch or 10mm wide metal earthling braid or #10 AWG /
6mm² insulated stranded wire with properly crimped ring terminals where required.
Heavier gauge conductors may be used. This vacuum system ground must be the
common point of all ground connections. Run grounding connections from here to the
Controller and Filament Transformer Box as well. In rack mount systems run the
grounding for the Controller the short way directly to the HVPS ground stud.
If no common grounding point is available on the vacuum system, the ground stud of the
HVPS may be used as common ground instead.
See section 7, Figures 7-A through 7-E.
3. Connect the supplied HV Output cable (RG213 Coax) between J5
HV OUT on the HVPS and the mating connector on the supplied Filament Transformer
Box. It is good practice to keep the HV cable away from sharp objects, machinery, and
high traffic areas. This will minimize the possibility of damage to the cable.
4. Connect the supplied Hirschmann cable between J3 FIL(AMENT) XFMR on the HVPS
and the mating connector on the Filament Transformer Box.
5. If desired, connect an external EMO (Emergency Off) switch with the supplied 2-pin
Hirschmann connector to J7 EXT EMO on the HVPS (not available on TT-3/6). Jumper
JP1 on the HVPS main board must be set to OFF (2-3) for activation of the external EMO
(contact Telemark Service for details)
6. Connect the supplied 25pin Sub-D M-F cable between J1 HV CONTROL on the HVPS
and J1 on the Controller.
7. Connect the supplied 9pin Sub-D M-F cable between J2 SOURCE CTRL on the HVPS
and J2 on the Controller.
8.
TT-3/6/8:
With input AC line voltage
OFF,
terminate the AC line feed wires to the supplied terminals
(P/N 124-1440-1), connector shell (P/N 124-1441-1), and cable clamp (P/N 124-1442-1).
It is mandatory to use 600V insulated stranded wire of sufficient gauge (min. #12AWG /
Содержание CHEETAH TT-10 CE 208 VAC
Страница 33: ...Telemark TT Cheetah Series Power Supply Instruction Manual 33 Figure 8 A Connections TT 3...
Страница 34: ...Telemark TT Cheetah Series Power Supply Instruction Manual 34 Figure 8 B Connections TT 6 8 208VAC Version...
Страница 35: ...Telemark TT Cheetah Series Power Supply Instruction Manual 35 Figure 8 C Connections TT 6 8 400VAC Version...
Страница 36: ...Telemark TT Cheetah Series Power Supply Instruction Manual 36 Figure 8 D Connections TT 10 15 208VAC Version...
Страница 37: ...Telemark TT Cheetah Series Power Supply Instruction Manual 37 Figure 8 E Connections TT 10 15 20 400VAC Version...
Страница 38: ...Telemark TT Cheetah Series Power Supply Instruction Manual 38 Figure 8 F J5 Remote User I O Connector...