B-1
TIENet
®
Model 350 Area Velocity Sensor
Appendix B Velocity Error Codes
B.1 Introduction
Erroneous flow data can result from a number of factors. The
area velocity system provides numbered error codes associated
with the 350 Velocity data to assist in troubleshooting.
The error codes are viewable using Isco Flowlink® software.
Definitions of the error codes are provided in Table B-1. For
further assistance, contact the factory.
B.2 Importing Data Dump
(.ddp) Files
Flow data from the Signature Flow Meter can be downloaded
onto a USB flash drive in the form of a .ddp (Data Dump) file.
To download the data:
1. Connect a flash drive to the USB port on the front panel of
the Signature. From the USB Options menu, select
Retrieve Data.
2.
Select “All data,” or specify a start date or date range, and
press NEXT. The data will be stored on the connected flash
drive in a folder called “ISCO.”
3. Connect the flash drive to a computer running Flowlink.
4. In Flowlink, select File > Import. When the import window
appears, browse to the folder containing the desired .ddp
file. Select the file and click Open.
Figure B-1 Signature flow data: Selecting the .ddp file(s)