TIENet
®
Model 350 Area Velocity Sensor
Section 2 Installation
2-8
5. Insert the reference tubing into the REF AIR port on the
case board, pushing it down inside the silicon tubing.
B
e
careful not to kink the reference tubing.
Figure 2-9 Insert the cable reference tubing into the case
board reference port
6. Gently tug the cable downward, to remove any slack within
the enclosure, taking care not to put any stress on the con-
nections.
7. Tighten the cord grip sealing nut (Figure 2-10).
Figure 2-10 Position and secure the cable
8. Close the front panel and fasten it shut with the two Phil-
lips screws.
CAUTION
If you are using conduit instead of the cord-grip fitting, the con-
duit must be sealed to prevent harmful gases and moisture
from entering the Signature enclosure. Failure to seal conduit
could reduce equipment life.
------------------------------------
Note:
Remove the red cap from
the reference port, on the case
board next to the TIENet connector,
being used by the 350 sensor