TIENet
®
301 pH/Temperature Device
Section 2 Installation and Setup for Signature
2-2
2. If using a cord-grip fitting, install the cable nut in the
appropriate opening on the bottom of the Signature enclo-
sure, securing it to the wall with the lock nut (concave side
facing wall).
3. Feed the TIENet device cable end through the sealing nut
and seal, and through the cable nut. Lightly tighten the
sealing nut, just enough to hold the cable in place while
installing the connector.
Figure 2-2 Installing cable with a cord-grip fitting
4. Attach the wire ends to the terminal strip as shown in
Figure 2-3, then press the terminal strip back down into its
socket on the case board, as shown in Figure 2-4, taking
care not to strain any wire connections. Gently tug each
wire when finished, to verify secure connection to the
screw terminals.
Note
The SHIELD wire is the bare drain emerging from the foil
s hi eld a r oun d th e Y E LLO W an d B R OW N w ir e s . T he
BRAID-DRAIN wire is the bare drain emerging from the sur-
rounding braided shield inside the cable jacket. It is not neces-
sary to prevent the two braids from coming into contact with
each other.
Lock Nut
(concave side
facing wall)
Cable Nut
Seal
(color may vary)
Sealing Nut