2-1

TIENet

®

 301 pH/Temperature Device

Section 2 Installation and Setup for Signature

2.1 Installation

External TIENet devices such as the 301 are all electrically con-

nected to the Signature flow meter in the same manner, usually

using conduit or cord-grip cable fittings. Multiple external

TIENet devices can be connected simultaneously.

2.1.1 Connecting the Cable

Refer to your Signature flow meter manual for instructions on

accessing the instrument’s interior components.

WARNING

Before proceeding, ensure that the flow meter has been 

disconnected from mains power.

Note

The steps that follow include instructions for installing cord-grip
fittings. Some applications will use user-supplied 

3

/

4

" ID con-

duit for cable routing.

1. Remove one of the 6-position plug-in terminal strip connec-

tors from the case board.

Figure 2-1 TIENet Device terminal strips

Содержание TIENet 301

Страница 1: ...TIENet 301 pH Temperature Device Installation and Operation Guide Manual Body 69 4303 071 Copyright 2021 All rights reserved Teledyne ISCO Revision E May 2021 ...

Страница 2: ......

Страница 3: ...dise Authori zation RMA specified Be sure to include a note describing the malfunction This will aid in the prompt repair and return of the equipment Teledyne Isco welcomes suggestions that would improve the information presented in this manual or enhance the operation of the equipment itself Teledyne Isco is continually improving its products and reserves the right to change product specification...

Страница 4: ......

Страница 5: ...ent d une manière autre que celles qui sont specifiées dans ce manuel la protection fournie de l instrument peut être affaiblie cela augmentera votre risque de blessure Hazard Severity Levels This manual applies Hazard Severity Levels to the safety alerts These three levels are described in the sample alerts below CAUTION Cautions identify a potential hazard which if not avoided may result in mino...

Страница 6: ...mboles de sécurité Ce symbole signale l existence d instructions importantes relatives au produit dans ce manuel Ce symbole signale la présence d un danger d électocution Warnungen und Vorsichtshinweise Das Ausrufezeichen in Dreieck ist ein Warnzeichen das Sie darauf aufmerksam macht daß wichtige Anleitungen zu diesem Handbuch gehören Der gepfeilte Blitz im Dreieck ist ein Warnzeichen das Sei vor ...

Страница 7: ... 4 pH Probe Calibration 2 10 2 4 1 Frequency 2 10 2 4 2 Procedure 2 10 2 5 Firmware Updates 2 12 2 6 Probe Storage and Maintenance 2 12 2 7 Contact Teledyne Isco 2 12 Appendix A Replacement Parts A 1 Replacement Parts A 1 A 1 1 301 pH Temperature A 2 List of Figures 1 1 Signature Flow Meter w Model 301 TIENet device and pH probe 1 1 2 1 TIENet Device terminal strips 2 1 2 2 Installing cable with a...

Страница 8: ...TIENet 301 pH Temperature Device Table of Contents vi ...

Страница 9: ... or 23m cable that attaches to the flow meter Attached to the device is a stainless steel submersible probe with combi nation type electrodes The probe has a porous PTFE liquid junction to resist fouling and coating and a steam sterilized glass hemi bulb for long term stability The probe has a standard 25 foot cable and built in exposed tem perature sensor Figure 1 1 Signature Flow Meter w Model 3...

Страница 10: ...nfold increase in H ion Each number between 7 and 14 equals a tenfold decrease of H ion pH measurements of wastewater are commonly made to monitor the effect of treatment chemicals added to raise or lower the pH Water that has been used for various industrial processes may deviate substantially from 7 Chemicals are often added to the water to bring the pH close to that of neutral water which is 7 ...

Страница 11: ...tion The potential when read by a sensitive voltmeter translates into a reading of pH With the Signature flow meter the voltage is sent first to a preamplifier inside the probe to reduce the impedance of the circuit and improve the signal to noise ratio and then on to the 301 to allow greater operating distance from the flow meter Note that pH probes are a consumable item and not covered under the...

Страница 12: ...301 device 301 pH Temperature Device with Signature connection ending in TIENet plug For use with portable Signature TIENet receptacle Includes combination pH probe with built in exposed temperature probe and 25 ft probe cable and one package of each buffer and rinse solution for probe calibration 10 m cable 60 4307 070 23 m cable 60 4307 071 Cut to length 60 4307 072 Cable lengths from Signature ...

Страница 13: ...multaneously 2 1 1 Connecting the Cable Refer to your Signature flow meter manual for instructions on accessing the instrument s interior components WARNING Before proceeding ensure that the flow meter has been disconnected from mains power Note The steps that follow include instructions for installing cord grip fittings Some applications will use user supplied 3 4 ID con duit for cable routing 1 ...

Страница 14: ...with a cord grip fitting 4 Attach the wire ends to the terminal strip as shown in Figure 2 3 then press the terminal strip back down into its socket on the case board as shown in Figure 2 4 taking care not to strain any wire connections Gently tug each wire when finished to verify secure connection to the screw terminals Note The SHIELD wire is the bare drain emerging from the foil shield around t...

Страница 15: ...Net 301 pH Temperature Device Section 2 Installation and Setup for Signature 2 3 Figure 2 3 TIENet Device terminal connections Figure 2 4 Attach wired terminal strip to case board socket Braid Drain Shield ...

Страница 16: ...closure taking care not to put any stress on the con nection 6 Tighten the cord grip sealing nut 7 Close the front panel and fasten it shut with the two Phil lips screws CAUTION If you are using conduit instead of the cord grip fitting the con duit must be sealed to prevent harmful gases and moisture from entering the Signature enclosure Failure to seal conduit could reduce equipment life ...

Страница 17: ...he connectors and push together Figure 2 6 2 The sensor release will click when the sensor connector is fully seated 3 Connect the two caps together 4 After the physical connection is made a scan must be per formed for the device to be recognized For additional TIENet connections use the TIENet Y cable or alternately an Expansion Box O Ring and Lubrication for the TIENet receptacle 1 Coat the O ri...

Страница 18: ...ing is more secure if the probe is facing upstream due to the stop tabs on the back end Remember to unscrew the plastic guard cap from the sensing end of the probe when you install it or the probe will be unable to sense the flow stream Installation in streams with high grease content will result in poor performance and require frequent cleaning and recali bration Greasy substances being nonconduc...

Страница 19: ...e probe in the ring or strap route the cable out of the stream so it will not trap debris or clog the sewer Note When installing the pH probe and its sensor carrier make sure the mounting slots on the carrier are completely pressed into the mating tabs on the ring The probe relies on a full engage ment between tabs and slots for secure mounting If the slots are loose against the tabs the probe may...

Страница 20: ...rom the Hardware Setup menu Configure refers to defining and selecting the parameters for each connected device The default parameters that will appear for the 301 device are 301 pH pH reading 301 Temperature Degrees C The name of any parameter can be changed by highlighting it and pressing Enter to display the character grid Nav igate the grid using the arrow keys Select characters with Enter and...

Страница 21: ...ration The sensors are being configured Please wait TIENet Setup Smart Sensor The sensors have been configured 301 pH 301 Temperature 2 301 Hardware Setup 1 Smart Sensor Setup TIENet 2 SDI 12 Setup 3 MODBUS Input Setup 4 MODBUS Output Setup 5 Modem Setup Press Enter for a list of sensors Scroll to the 301 and press Enter to select Scroll with arrow keys to highlight select deselect any displayed p...

Страница 22: ...H Consider replacing the probe when delays become apparent 2 4 pH Probe Calibration The Signature flow meter provides a calibration procedure for the pH probes using commercially prepared buffer solutions 2 4 1 Frequency For accurate readings you must clean and re calibrate the probe on a regular basis Calibration intervals typically range from weekly to monthly depending on the application Note F...

Страница 23: ...s Select NEXT to continue Adjust Options Level pH 301 pH Calibration 301 pH X XX 1 2 or 3 4 1 Site Setup 2 Measurement Setup 3 Adjust 4 Equation Trigger Setup 5 Data Storage Push Setup 6 Sampler Setup 7 Inputs Outputs Alarms Setup 8 Reset Totalizers 9 Reports History Setup If the pH buffer is not the exact value enter the value printed on the side of the container The screen will cycle repeatedly ...

Страница 24: ...f a pH probe in a 4 0 buffer solution is recommended as this has a regenerative effect on the glass and does not put a memory on it Tap water will work if 4 0 buffer solution is not available Deionized water is good for quick rinses to clean the element but not for prolonged storage of an electrode Con tinuous exposure of the ion sensitive membrane to a wetted but non ionic solution will improperl...

Страница 25: ...strations Refer to the call out in the adjacent table to determine the part number for the item Replacement parts can be purchased by contacting Teledyne ISCO s Customer Service Department Teledyne ISCO Customer Service Department P O Box 82531 Lincoln NE 68501 USA Phone 800 228 4373 402 464 0231 FAX 402 465 3022 E mail IscoInfo teledyne com ...

Страница 26: ...TIENet 301 pH Temperature Device A 2 A 1 1 301 pH Temperature REPLACEMENT PLUG SENSOR CAP KIT FOR SKU 609004469 2100 SERIES AND TIENET DEVICES ...

Страница 27: ...n the product O 嫷䯉年㦘㹡㦘 䓸德 年捷ↅ㓏㦘 德㧟㠨 䤓 摞 ST 㪖 屓 䤓棟摞尐㻑ⅴₚᇭ O Represent the concentration of the hazardous substance in this component s any homogeneous pieces is lower than the ST standard limitation X 嫷䯉年㦘㹡㦘 䓸德咂 年捷ↅ䤓㩟 德㧟㠨 䤓 摞怔 ST 㪖 屓 䤓棟摞尐㻑ᇭ 㷳 㫈㗽 棔 ₙ嫷 㓢 X 䤓㔏㦾 扪嫛扪 㷴広㢝ᇭ X Represent the concentration of the hazardous substance in this component s at least one homogeneous piece is higher than the ST stan...

Страница 28: ......

Страница 29: ... if the instrument proves to be defective within 30 days of shipment Throughout the remainder of the warranty period the customer will pay to return the instrument to Teledyne Isco and Teledyne Isco will pay surface transportation to return the repaired instrument to the customer Teledyne Isco will not pay air freight or customer s packing and crating charges This warranty does not cover loss dama...

Страница 30: ......

Отзывы: