50
To Measure AC Current
Range:
200μA, 2 mA, 20 mA, 200 mA, 2A, 10A
Max Resolution:
0.1nA (in the range of 200μA)
Input Protection:
10A protection is available in all ranges and a 10%
over-range for all ranges except the 10A range. If the reading exceeds
the range, “
overload
” will be displayed.
Operating Steps:
1. Enable the ACI measurement
Press
and
on the front panel to enter the AC current
measurement interface, as shown in Diagram 2- 12.
Diagram 2- 12 AC Current Measurement Interface
2. Make the connection
Connect the test leads with the measured signal by referring to
“
Measurement Connections
”.
3. Set the range
Press
【
Range
】
to select a range for the measurement. You can also
use the
,
, and
keys on the front panel to select the range.
Auto (autorange) automatically selects the range for the measurement
based on the input. Autoranging is convenient, but it results in slower
measurements than using a manual range. Autoranging goes up a range
at 110% of the present range, and down a range when below 10% of the
present range.