6.Start scan operation
Set the [Start] to on to start the scan operation
The upper part of the interface displays the function, range, and result of the current channel. The table below records the measurement result of each channel.
7.Enter trend chart and statistics mode (optional operation)
Press
and
to open trend chart and statistics function.
Set the [Inquire Channel] by direction keys and the interface displays the minimum, average, maximum, span, standard deviation, samples and trend chart of
measurement results of the setting channel during the scan operation.
Press [Done] to return to higher level menu.
8.Stop scan operation
Set the [Start] to off to stop the scan operation.
If the [Cycles] is set to manual, then the instrument will stop scan when it reaches the cycle number
9.Store measurement data (optional operation)
Press
and
, then Select [Store/Recall]
→
[Store Settings] to enter the storage function interface.