MERIDIAN
UNIVERSAL GAS DETECTOR
CONFIGURATION & SETUP
5-18
USE AND DISCLOSURE OF DATA
Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is
prohibited.
5.1.5 Configuring the Datalog Menu
Use the magnetic keys to navigate through the menu structure as necessary (refer to
).
Figure 5-6: Datalog Menu Structure
Table 5-7: Datalog Menu Parameters
Item
Sub-Item
Description*
DataLog
Show Log*
Displays time stamped events. Events include: When device went in and out of
Alarms and when device went in and out of Faults. YY MM DD: Year, month
and day of the log. Each event line format is: HHMM: Hour and minutes from
the date stamp, EC: Event Code from the logged event and EventData.
Event codes
EC EVENTDATA= DESCRIPTION
SU YYMMDDHHMMSS=Startup
SD YYMMDD=Set date, with YMD
ST HHMMSS=Set time, with HMS
FC Low Volts=Brownouts below 10V
FC SV Mismatch=SV Mismatch
FC XXXXXXXXWDog=SW WDT Fire
FC XXXXXXXXDiag=Diag Safety
FC MB_Address=MB Address Changed
FC >1 HP Sensor=More than one high power sensor
FC XXXXh SnF=Sensor fault
LU Sn=Linkup, with S (Sensor number)
Ax sss.s:rrr.s=Alarm set point/reset point (right after LU event)
LD Sn=Linkdown, with S (Sensor number)
IO Sn=Inhibit on, with S (Sensor number) or SYS
>
>
>
>
>
>
MAIN MENU
INFORMATION
SETUP
CALIBRATION
DATALOG
DISPLAY
USER ACCESS
SHOW LOG
CLEAR LOG
DATALOG MENU
>
>
______________________
0245 PW 0 ACCEPTED
0246 A1 OS110.100
0248 A1 FS18.9000
0250 PW 0 FAILED
12 09 26 EVENT LOG