Operator's Manual
4. Press each button and knob on the Front Panel in turn. As you press them, you should see the cor-
responding block on the simulation turn blue, indicating proper functioning. If any block fails to change, the
control is malfunctioning.
5. Press the number 8 on the keypad three times to exit the test.
LED Test
This test determines whether the LEDs that light the Function buttons on the Front Panel buttons are working
properly.
1. Press the
Utility
function button.
2. Press the
More
softkey, then press
Test/Cal
.
3. Press
SelfTest
, then
LEDTest
. The display changes to show a simulation of the Front Panel exactly like the
one in the Key Test.
4. Continue to press the
number 7
on the keypad. Each block on the simulation that corresponds to a Func-
tion button turns blue, while the button itself lights up, indicating proper functioning. If any block or button
fails to light, the control is malfunctioning.
5. Press the number 8 on the keypad three times to exit the test.
Self Calibration
WaveStation includes a self-calibration routine that makes gain and phase adjustments. Run the self
calibration on an as needed basis.
Contact your nearest
Teledyne LeCroy service center
to schedule a factory calibration once per year.
1. Press the
Utility
function button.
2. Press the
More
softkey, then press
Test/Cal
.
3. Press
SelfAdjust
. You will see a progress bar appear over the display. Wait until the calibration completes,
then press any Function button to exit.
922869 Rev B
43