![Teledyne Foxy R1 Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/teledyne/foxy-r1/foxy-r1_instruction-manual_1080518077.webp)
Foxy
®
R1 and Foxy
®
R2 Fraction Collectors
Section 4 Serial Command Control
4-17
4.5.4 Problem Responses
The fraction collector will respond with a
PROBLEM=<type>
message should the command fail. The following types may be
reported.
IMPROPER MODE
The program values cannot be changed
while a run is in progress.
Note
RUN and STOP cannot be executed while in manual control.
INVALID COMMAND
One of three situations could cause
this response:
• An unknown command has been entered, usually a
typographical error.
• An operand has been sent for a command that does not
require one.
• An operand is missing for a command that requires one.
INVALID OPERAND
The operand value is out of range or
unsupported.
LOCAL MODE
The unit must be in the remote mode to do
anything other than read values.
Содержание Foxy R1
Страница 2: ......
Страница 4: ......
Страница 12: ...Foxy R1 and Foxy R2 Fraction Collectors Safety xii ...
Страница 17: ...Foxy R1 and Foxy R2 Fraction Collectors Section 1 Introduction 1 5 Figure 1 2 Fraction collector features ...
Страница 24: ...Foxy R1 and Foxy R2 Fraction Collectors Section 1 Introduction 1 12 ...
Страница 38: ...Foxy R1 and Foxy R2 Fraction Collectors Section 2 Preparation for Use 2 14 ...
Страница 60: ...Foxy R1 and Foxy R2 Fraction Collectors Section 3 Programming 3 22 ...
Страница 78: ...Foxy R1 and Foxy R2 Fraction Collectors Section 4 Serial Command Control 4 18 ...
Страница 92: ...Foxy R1 and Foxy R2 Fraction Collectors Appendix A Icon Glossary and Menu Charts A 10 ...
Страница 96: ...Foxy R1 and Foxy R2 Fraction Collectors Index Index 4 ...
Страница 98: ......
Страница 100: ......
Страница 102: ......