Foxy
®
R1 and Foxy
®
R2 Fraction Collectors
Section 2 Preparation for Use
2-8
Figure 2-9
High flow diverter valve ports (front view)
CAUTION
Risk of fire or equipment damage caused by electrostatic dis-
charge. Organic solvents delivered at high flow rates can build
an electrostatic charge. The conductive black tubing supplied
in the kit dissipates the electrostatic charge. If handling organic
solvents at high flow rates, do not substitute the conductive
tubing.
Note
Cut the tubing with a tubing cutter to ensure that the end is cut
cleanly and free from any deformation.
4. Prepare one end of each tubing piece as shown in Figure
Figure 2-10 Tubing preparation, high flow
5. Using two wrenches, connect the In and Waste tubing to
the valve using caution to prevent damage to the plastic
valve body.
Waste
(NO)
IN
(Common)
Drop Former
(NC)
Fr
o
n
t Fe
rr
ul
e
Nut
Ba
c
k
Fe
rr
ul
e
V
alve Fittin
g
Tu
b
in
g
Содержание Foxy R1
Страница 2: ......
Страница 4: ......
Страница 12: ...Foxy R1 and Foxy R2 Fraction Collectors Safety xii ...
Страница 17: ...Foxy R1 and Foxy R2 Fraction Collectors Section 1 Introduction 1 5 Figure 1 2 Fraction collector features ...
Страница 24: ...Foxy R1 and Foxy R2 Fraction Collectors Section 1 Introduction 1 12 ...
Страница 38: ...Foxy R1 and Foxy R2 Fraction Collectors Section 2 Preparation for Use 2 14 ...
Страница 60: ...Foxy R1 and Foxy R2 Fraction Collectors Section 3 Programming 3 22 ...
Страница 78: ...Foxy R1 and Foxy R2 Fraction Collectors Section 4 Serial Command Control 4 18 ...
Страница 92: ...Foxy R1 and Foxy R2 Fraction Collectors Appendix A Icon Glossary and Menu Charts A 10 ...
Страница 96: ...Foxy R1 and Foxy R2 Fraction Collectors Index Index 4 ...
Страница 98: ......
Страница 100: ......
Страница 102: ......